background image

8

This table is intended to assist in resolving user-correctable Refrigerator problems by relating symptoms to
their likely causes. If service beyond the scope of this table is required, contact your nearest Fisher Scientific
Service Office.

Troubleshooting

Symptom 

Probable Cause  

Action 

Does Not Run  

Unit Unplugged 

Plug in Unit 

Blown fuse or tripped circuit 

Check fuse or circuit breaker at

breaker breaker 

box 

Runs Continuously  

Frost buildup on refrigeration coils 

Defrost unit

Clicking Sound  

The compressor is equipped 

Disconnect power. Allow refrigeration

with a thermal protector.

system time to equalize, then try 

This device shuts off the

again. If problem persists, call for

compressor when it

service.

becomes too hot. A clicking
sound occurring about    
every 30 seconds indicates 
this protector is working

Insufficient Cooling  

Thermostat set too high 

Reduce thermostat setting 

Unit frosted 

Defrost unit

Содержание Isotemp 150

Страница 1: ...Isotemp Explosion Proof Refrigerators and Freezers Refrigerator Model 153 153HK Freezer Model 150 150HK 104354 9 15 06...

Страница 2: ...5 Leveling The Unit 5 Shelves 5 Electrical Connection 5 Operation 6 Initial Startup 6 Temperature Control 6 Condensate Disposal 6 Material Compatibility 7 Troubleshooting 8 Maintenance 9 Cabinet Clea...

Страница 3: ...lication alterations or modifications will void the Warranty This apparatus is designed for use in Class I II or III loca tions as defined by the National Electrical Code unless otherwise noted Safety...

Страница 4: ...ibe such exter nal evidence of loss or damage may result in the carrier s refusing to honor your damage claim The form required to file such a claim will be supplied by the carrier Concealed Loss or D...

Страница 5: ...if it is left halfway open Shelves Shipped inside each cabinet are four fixed shelves which also serve as part of the evaporator coil Electrical Connection Determine the total amount of current prese...

Страница 6: ...Remove the two screws with a flat blade screw driver for access For operation above an altitude of 3000 ft have the thermo stat adjusted by a technician Condensate Disposal The condensate drain is at...

Страница 7: ...d by D A T A and The International Plastics Selector Inc Cordura Companies 9889 Willow Creek Road P O Box 26637 San Diego California 92126 Aluminum Alloys are susceptible to corrosion when exposed to...

Страница 8: ...rcuit Check fuse or circuit breaker at breaker breaker box Runs Continuously Frost buildup on refrigeration coils Defrost unit Clicking Sound The compressor is equipped Disconnect power Allow refriger...

Страница 9: ...of water is recommended for cleaning the interior and exterior of the cabinet All surfaces should be rinsed and thoroughly dried Condenser The condenser coil is located behind the sides and top exter...

Страница 10: ...C 4 C at 20 C 4 C Electrical Requirements 13 986 150 115 Volts 10 60 Hz 5 0 Amps 13 986 150HK 220 Volts 10 50 Hz 5 0 Amps 13 986 153 115 Volts 10 60 Hz 5 0 Amps 13 986 153HK 220 Volts 10 50 Hz 5 0 Am...

Страница 11: ...11...

Страница 12: ...adjustment is reasonable in the light of the manufacturer s own warranty b The Company shall be released from all obligations under its warranty in the event repairs or modifications are made by perso...

Отзывы: