6
Operation
Initial Startup
This unit is not equipped with a power switch. Turn the
applicable facility circuit breaker off during installation.
Simply turn power on the facility’s circuit that supplies the
unit.
Temperature Control
To adjust the interior temperature down, simply turn the
thermostat knob clockwise. The adjustment knob is locat-
ed behind the front cover of the blue temperature control
unit that is on the top exterior on the cabinet. Remove the
two screws with a flat blade screw driver for access. For
operation above an altitude of 3000 ft., have the thermo-
stat adjusted by a technician.
Condensate Disposal
The condensate drain is at the center of the interior floor.
The condensate drain tube is connected to this drain and
is attached to the left front bottom of the unit behind the
front sheet metal. Unhook and place the drain tube into a
suitable pan and remove the drain plug. Turn the temper-
ature control up or remove power to the unit and allow the
unit to defrost.
Warning
If the unit is tilted in excess of 30
degrees, do not apply electrical power
for a minimum of 12 hours.