Isotemp ULT Chest 6-3
Fisher Scientific
Section 6
Factory Installed Options
3. Carefully remove the existing Permagum sealant from around the
probe port opening.
4. Open the freezer lid and locate the probe cover on the upper front left
wall. Remove the two Phillips head screws securing the probe cover
(see Figure 6-5).
5. Route the BUS probe through the probe port, approximately 12”.
Secure the back-up probe to the temperature probe using a small tie
wrap (Figure 6-5).
6. Seal around the interior and exterior opening of the probe port with
Permagum sealant.
7. Reinstall the probe cover (Figure 6-5).
micro
board
battery
bracket
battery
switch
probe
access
port
Figure 6-4.
Probe Access Port
Figure 6-5.
Probe location
Install Temperature Probe
(continued)