The following section describes the set up and operation of the optional
chart recorder.
1. Open the plastic door of the recorder and press button #3 until the
pen begins to move outward.
2. Unscrew the knob at the center of the chart and remove the paper.
3. Install the new chart paper, position the paper to the correct time line
and replace the knob.
4. Remove the cap from the felt pen and press button #3.
The chart recorder contains eight temperature ranges and is factory-
programmed for the freezer. To change the recorder range:
1. Press and hold button #3 for one second, then let the pen move off the
chart paper.
2. Press and hold for five seconds either button #1 or button #2.
3. Release the button and the green LED will
begin to flash. Count the number of flashes
to determine the present program setting.
4. To change the program setting, press the
left or right arrows to increase or decrease
the count.
5. When the desired program number is
flashing, press button #3 to bring the pen
arm back onto the chart. Recording will
begin in the new program.
The recorder must be in service for 24 hours before performing the
following calibration procedure.
1. Place an accurate thermometer in the chamber next to the recorder
probe.
2. Temperature probes for the recorder are located in the left front corner
of the freezer chamber (Figure 1-4).
3. After about three minutes, compare the thermometer reading with the
chart recorder reading.
Isotemp ULT Chest 6-7
Fisher Scientific
Section 6
Factory Installed Options
Isotemp ULT Chest 6-7
Fisher Scientific
Install Chart Paper
Change Program
Program
From
To
1
-40
30°C
2
0
60°C
3
-100
38°C
4
-5
50°C
5
0
100°C
6
-100
200°C
7
-115
50°C
8
-10
70°C
Calibrate Chart Recorder