Power
- indicates the unit has AC power.
Low Battery
- battery charge is low. The battery needs replaced or
recharged.
Solenoid Engaged
- BUS has opened the solenoid so it can inject gas (CO
2
or LN
2
).
Press-To-Test
- Activates the solenoid and injects LN
2
or CO
2
into the
freezer chamber as long as the button is depressed. The solenoid engaged
indicator should light. If the Low Battery indicator lights during the test,
replace the BUS battery.
Note
The solenoid will not engage if lid is open.
s
The optional BUS can be configured for LN
2
or CO
2
supply. Section 3 -
Configuration contains instructions for setting the BUS type.
The optional back up system is designed to inject CO2 or LN2 into the
freezer compartment if the temperature rises above back up system set
point. To set the BUS set point:
1. Press the Mode key until the Settings indicator lights.
2. Press right arrow until “BACKUP = -XX” displays in message center.
3. Press up or down arrow key until desired BUS set point is displayed.
4. Press Enter to save the setting.
5. Press the Mode key until the Run indicator lights, or press the right or
left arrow to go to the next or previous parameter.
If no control keys are pressed, the freezer will automatically return to
RUN mode after 5 minutes.
Isotemp ULT Chest 6-5
Fisher Scientific
Section 6
Factory Installed Options
Configure Optional BUS
(Back Up System)
Figure 6-6.
BUS Control Panel
BUS Control Panel
(continued)
Set Optional BUS Set
Point