Locate the freezer on a firm, level surface in an area with an ambient
temperature between 18°C and 32°C. Provide ample room to reach the
mains disconnect switch (power switch) located on the rear of the freezer.
Note
For proper ventilation and airflow, a minimum clearance of 5” at the
rear and front and a clearance of 8” on the side of the freezer is required.
Allow adequate space for lid opening. If ambient increases above 36°C,
clearance at the rear of the cabinet must be increased to 8”.
s
The parts bag, located inside the cabinet, contains the following parts.
Install the bolts into the pre-tapped holes on the back of the compressor
section. Install a neoprene cap on each bolt. Refer to Figure 1-2 for the
locations of the pre-tapped holes.
Isotemp ULT freezers have a data communications interface. The factory
default setting is RS-232.
The wiring identification for the interface is
shown in Figure 1-10. One nine pin, sub
"D" style connector is located on the back
of the freezer. See Figure 1-2 for the location
of the connector on the freezer.
The freezer transmits temperature information every 60 minutes. A
standard DB9 serial extension cable can be used to connect the freezer to a
serial device. Some serial devices may require a null modem adapter.
Data format:
Baud . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1200
Data bits . . . . . .8 (7 bit ASCII with leading zero)
Start bits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Stop bits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
Parity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .none
1-8
Isotemp ULT Chest
Fisher Scientific
Section 1
Installation and Start-up
Quantity
Stock #
Description
Purpose
2
510016
1/4-20 x 5-1/2” Bolt
Wall Bumper
2
380520
Neoprene Cap
Cap Protector
Choose Location
Install Wall Bumpers
RS-232 Communications