FISCHER DELTASCOPE FMP10 Скачать руководство пользователя страница 6

Brief Guide FMP10 / FMP20

Page 3

Description of the Instrument

3

Description of the Instrument

Fig. 3-1 Front and rear view of the instrument, connections

1

Probe connector socket, 

 Page 8

2

LCD Display, 

 Page 4

3

Keys for directly retrieving 
functions, 

 Page 5

4

USB port for connecting a 
PC

5

ON/OFF key to turn the in-
strument on or off, 

 Page 9

6

Cover; additional function 
keys can be found under the 
cover, 

 Page 5

7

Non-slip rubber supports

8

Foldable instrument stand

9

Battery compartment, 

 Page 7

1

2

3

4
5

6

7

8

9

7

Содержание DELTASCOPE FMP10

Страница 1: ...DELTASCOPE FMP10 ISOSCOPE FMP10 DUALSCOPE FMP20 Guidelines Coating Thickness Material Testing Microhardness Material Analysis...

Страница 2: ......

Страница 3: ...Install the batteries rechargeable batteries 7 6 2 Probe connection 8 6 3 Instrument On Off 9 7 Adjusting Instrument and Probe 11 7 1 Assigning a New Probe 11 7 2 Normalization 12 7 3 Corrective Calib...

Страница 4: ...tensive oper ator s manual can be found on the CD ROM that is supplied with the instrument The following symbols are used in this brief guide Indicates safety information referring to danger for per s...

Страница 5: ...non conducting non magnetic coatings on non ferro magnetic electrically conducting base materials Paint lacquer or plastic coatings on for example aluminum copper zinc etc as well as anodized coating...

Страница 6: ...obe connector socket Page 8 2 LCD Display Page 4 3 Keys for directly retrieving functions Page 5 4 USB port for connecting a PC 5 ON OFF key to turn the in strument on or off Page 9 6 Cover additional...

Страница 7: ...rement uses the eddy current method Padlock Restricted operating mode is enabled i e the keys ZERO and CAL are not active the service functions cannot be retrieved Arrow circle Free running display is...

Страница 8: ...rieves the corrective calibration z Changes the information shown on the LCD display z The display will change faster if is pressed for longer than 3 seconds z Changes the information shown on the LCD...

Страница 9: ...easuring beginning on Page 16 z Always hold the probe at its grip sleeve right figure z Always place the probe gently and at a right angle on the specimen surface z Slide the grip sleeve to the specim...

Страница 10: ...AM3 MN1500 batteries or 4 individual rechargeable batteries 1 2 V 2400 mAh Type AA Using other batteries may lead to instrument damage Use only non damaged batteries rechargeable batteries If the bat...

Страница 11: ...ON OFF key on the right side of the control panel The LCD display is not backlit and no characters are visi ble Protect the instrument and accessories from electrostatic charges Electrical discharges...

Страница 12: ...s of the LCD display will appear briefly at the same time Page 4 At the end of the monitoring routine the instrument is ready to make measurements The icon of the measure ment method of the connected...

Страница 13: ...e connected NF Fe NC NF Dual probe connected and dual method both set i e in the open application both measurement methods can be used to make measurements NF Fe NC NF Dual probe connected and dual me...

Страница 14: ...006 Probe changed The probe must be assigned Chapter 7 1 7 1 Assigning a New Probe Key sequ Action Detail of the LCD display Explanation This warning appears briefly after power up if a probe other th...

Страница 15: ...n is per formed s Standard deviation n Number of measurements Base material Fe NF Measurements should be performed on an uncoated spec imen Fe or NF display for the type of base material When using du...

Страница 16: ...shed for the calibration curve and stored in the instrument Required materials z Reference part uncoated part from the production z Calibration Standards Selecting the Calibration Standards for the Co...

Страница 17: ...f base material will be displayed only after the 1st measurement is completed Perform measurements on the uncoated specimen base material normalization Make measurements at various points of the refer...

Страница 18: ...er formed on the calibration standard and then the rated value is corrected using the arrow keys This step is not required if as shown in the example the rated value for the thick ness of the calibrat...

Страница 19: ...ration respectively 8 Measuring To make a measurement place the probe at a right angle on the specimen surface The probe can be lifted off after the measurement acquisition i e after the read ing appe...

Страница 20: ...ument on After pressing ENTER 5 times the identifi cation number 157 appears on the LCD display 2 x Enter the identification number 159 by pressing the arrow key 2 times Use ENTER to confirm the setti...

Страница 21: ...ee running mode Instrument mode Restr Mode Analog display Measurement Audible signal Measurement effect External start Measuring mode Standard Area measurement Automatic measurement Display dual metho...

Страница 22: ...29 34 51 japan helmutfischer com www helmutfischer com FRANCE FISCHER INSTRUMENTATION ELECTRONIQUE Parc d Activit s Nord du Pas du Lac 5 rue Micha l Faraday F 78180 Montigny le Bretonneux BP 289 F 78...

Страница 23: ......

Отзывы: