Brief Guide FMP10 / FMP20
Page 11
Adjusting Instrument and Probe
7
Adjusting Instrument and Probe
The following options exist after the instrument is turned on:
1.
The instrument recognizes the connected probe. The measurement screen
appears on the LCD display. The display of the measurement method does
not flash. In this case, you can start immediately making measurements.
2.
If the instrument is turned on and a probe other than the one used last is
connected, the following display will appear: “W006 Probe changed”. The
probe must be assigned (
Chapter 7.1).
7.1
Assigning a New Probe
Key sequ. /
Action
Detail of the LCD
display
Explanation
This warning appears briefly after power-
up if a probe other than the last one used
is connected to the instrument.
After a brief time, the display to the left will
appear.
Example: FD10 = Name of the connected
probe
DEL: Probe assignment starts
ENTER: Probe will not be assigned, mea-
surement method display flashes
DEL
DEL: All stored readings will be deleted.
ENTER: The probe will not be assigned,
measurement method display continues to
flash
ZERO and CAL appear on the display. A
normalization and a corrective calibration
must be performed
(
beginning on
Page 12).