Page 14
Brief Guide FMP10 / FMP20
Adjusting Instrument and Probe
Procedure
Key sequ. /
Action
Detail of the LCD
display
Explanation
CAL
Use CAL to start the corrective calibration.
CAL appears and remains on the LCD dis-
play while the corrective calibration is per-
formed.
[s]: Standard deviation
[n]: Number of measurements
[base material (Fe/NF)]: Measurements
should be performed on an uncoated spec-
imen (Fe or NF display for the type of base
material), i.e., a Normalization is carried
out.
When using dual probes, the type of base
material will be displayed only after the 1st
measurement is completed.
Perform measurements on the uncoated
specimen (base material) (normalization).
Make measurements at various points of
the reference area.
The mean value of all readings obtained
for the normalization will be displayed.
ENTER
[Entry:
∧∨
]: Use the arrow keys to set the
nominal thickness value of the calibration
standard. The rated value is printed onto
the calibration standard.
The rated value can be set faster if a mea-
surement is performed on the calibration
standard and then the rated value is cor-
rected using the arrow keys.
[CAL Nom. v. 1: 23.70]: Display of the set
rated value for the thickness of the calibra-
tion standard (Example: 23.7 µm)