Brief Guide FMP10 / FMP20
Page 13
Adjusting Instrument and Probe
7.3
Corrective Calibration
Options
One-point calibration with one calibration standard:
A new zero point and
one additional point are established for the calibration curve and stored in the in-
strument.
Two-point calibration with two calibration standards:
A new zero point and
two additional points are established for the calibration curve and stored in the
instrument.
Required materials
z
Reference part (uncoated part from the production)
z
Calibration Standards
Selecting the Calibration Standards for the Corrective Calibration
A corrective calibration using the calibration standards included with the probe
shipment provides the best measuring precision for the entire measurement range
of the probe.
ENTER
The new characteristic will be computed
automatically and stored. The instrument
is again ready to make measurements.
Key sequ. /
Action
Detail of the LCD
display
Explanation