Brief Guide FMP10 / FMP20
Adjusting Instrument and Probe
7.2
Normalization
With the normalization, a new zero point is established for the calibration curve
and stored in the instrument.
Required materials
Reference part: Uncoated part from the production.
Procedure
Key sequ. /
Action
Detail of the LCD
display
Explanation
ZERO
Use ZERO to start the normalization.
ZERO appears and remains on the LCD
display while the normalization is per-
formed.
[s]: Standard deviation
[n]: Number of measurements
[Base material (Fe/NF)]: Measurements
should be performed on an uncoated spec-
imen (Fe or NF display for the type of base
material).
When using dual probes, the type of base
material will be displayed only after the 1st
measurement is completed.
Perform measurements on the uncoated
specimen (base material).
Make measurements at various points of
the reference area.
The mean value of all readings obtained
for the normalization will be displayed.
ENTER
A confirmation indicating that the normal-
ization has been carried out successfully
appears.
Pressing ENTER confirms the message.