6.12 STANDARD CHECK-WEIGHING MODE (WHEN F25 = MODE 1)
31
: -
This mode is used to compare the weight reading obtained with the preset
HI and LO limit set to this scale. The comparison result (HI, OK or LO) will
then be displayed with or without buzzer
32
.
If a relay module is installed, the comparison results are also sent through
the relay module.
Standard checkweighing mode can be activated during the below functions
mode. Bracketed are the targets to be checked under various functions.
Weighing (weight value),
Piece Count, (number of pieces),
Percentage (percentage value),
Animal Weighing (actual weight value
33
).
6.13 INFLOW/OUTFLOW CONTROL LOGIC
The inflow/outflow logic mode is built-in the software to simplify the control
system.
To achieve direct inflow/outflow control, LO output should be connected and
used control start/stop of the inflow device; while HI output should be
connected and used to start/stop of the outflow device.
Refer to
DIAGRAM: - INFLOW / OUTFLOW
for more information of this
control logic.
6.14 ABOUT NEAR ZERO VALUE (F26)
Nero zero value is useful for both dynamic weighing applications. It is used
to avoid false LO signal when a load is approaching and leaving the
weighing platform.
HI/OK/LO comparison will only start when weight reading exceeds the near
zero value entered. Refer to
DIAGRAM: -NEAR ZERO VALUE
for more
information.
31
Before using this function mode, set preferred near zero weight value in
F26
. System will
ignore any near zero value entered which is less than 20e. If it is the case, system will use
20e as minimum near zero weight value. For static weighing application, set near zero
value to zero.
32
Set F15 to obtain the required buzzer output configuration.
33
Despite the weight value is being held (frozen), this instrument will still detect continuously
the actual weight applied to the platform, compare this actual applied with the HI and LO
limits set and present the comparison result based on the actual weight applied.