background image

FERNO/MAN/1121/226-R3/UK

35

10  - MAINTENANCE

10.1 - Maintenance Schedule

The trolley requires regular maintenance. Set up and follow 
a maintenance schedule. The table at right/below represents 
minimum intervals for maintenance. Keep maintenance records. 
A sample maintenance record sheet is provided at the end of this 
manual.
When using maintenance products, follow the manufacturers’ 
directions and read the manufacturers’ material safety data sheets. 
You can purchase a recommended disinfectant from your Ferno 
distributor or Ferno UK Service Department.

IMPORTANT

Disinfectants and cleaners containing chlorine/bleach, 

phenol, or iodine can cause damage. It is recommended not 

to use products containing these chemicals. If these type of 

disinfectants are to be used, then to minimise the level of 

damage they will cause it is recommended that they are used 

with the least possible contact time, diluted as specified by 

the manufacturer’s guidelines, and that the equipment is 

then thoroughly washed down with water after disinfecting 

in order to remove any chemical residue and prevent any 

ongoing surface damage from occurring. Accessories such as 

push poles, IV poles etc. should be cleaned and disinfected 

whilst detached from the trolley and the fixing points on the 

trolley disinfected and thoroughly washed down after. 

Maintenance

10.2 - Cleaning the Trolley

Regular cleaning helps reduce the risk of transmitting disease and 
enables equipment to function at its optimum.
Ensure that all equipment is removed from the trolley and that 
the Olson Power distribution socket is disconnected from mains 
supply
Trolley should be hand-wiped using a suitable cleaning solution. 
For particularly grimy cases, especially if the trolley has been in 
storage for a long period, it may be necessary to first clean the 
trolley with a water-soluble solvent-soaked cloth.
Once cleaned, dry the trolley with a soft towel, paying particular 
attention to swivel and sliding joints. If a compressed air hose is 
available, you may find that compressed air jets are more effective 
in drying the hard-to-reach places..
Ensure no liquid enters the power distribution sockets.

10.3 - Disinfecting the Trolley

Wipe all surfaces with disinfectant. Follow the disinfectant 
manufacturer’s instructions for application method and contact 
time. Ferno recommends you inspect the trolley for damage as 
you clean and disinfect it. See note regarding disinfectants to use.
1. Remove the mattress, patient harness, straps and any 
accessories. 
2. Hand clean all surfaces of the trolley with warm water and a 

mild detergent. 

3. Rinse with warm, clear water. Dry the trolley with a towel 

and allow it to air-dry thoroughly, replacing the mattress and 
harness/straps prior to storage. 

4. To disinfect: Apply disinfectant to the trolley, following the 

disinfectant manufacturer’s instructions for application method 
and contact time. Ensure to wash down thoroughly after 
disinfecting to remove any chemical residue that might cause 
long-term damage

10.4 - Disinfecting & Cleaning Harness

It is recommended to remove the harness, straps and mattress 
from the trolley before cleaning and disinfecting to ensure they 
can be  cleaned effectively on all surfaces.
Cleaning Webbing straps
1. Immerse webbing straps in a solution of mild soap and water to 
clean Do not immerse any metal buckles or metal fixation clips in 
the solution, these should be wiped clean only.
2. Repeatedly dip the webbing in clean water to rinse. Do not 
immerse any metal buckles or fixation clips. Hang the straps to 
air-dry. Attach only clean, dry straps back onto to the equipment.
Disinfecting webbing straps
1. Immerse or spray the webbing straps following the selected 
disinfectant manufacturer’s instructions for application method 
and contact time. Do not immerse any metal buckles or metal 
fixation clips into disinfectant liquid
2. Spray the metal buckles and fixation clips with disinfectant 
cleaner spray, following the disinfectant manufacturer’s 
instructions for application method and contact time
3. Hang the straps to air-dry if needed. Attach only dried straps to 
the equipment
Cleaning Biosafe straps
1. Wipe-clean the straps, buckles and fixing clips with mild soap 
and water
2. Hang the straps to air-dry. Attach only clean, dry straps back 
onto to the equipment
Disinfecting Biosafe straps
1. Spray the webbing straps, metal buckles and fixation clips 
with disinfectant cleaner spray, following the disinfectant 
manufacturer’s instructions for application method and contact 
time
2. Hang the straps to air-dry if needed. Attach only dried straps to 
the equipment

 

WARNING

Improper maintenance can cause injury. Maintain the 

trolley only as described in this manual.

Minimum  

Maintenance Intervals

Each U

se

A

s N

eeded

Each M

on

th

Disinfecting 

Cleaning

 

Inspecting

 

Lubricating

 

Содержание CCT-PX

Страница 1: ...rs Manual FERNO MAN 1121 226 R3 UK Stock Code 2003 0158 Read this Manual and Retain for Future Reference Photo showing CCT PX configured with various available options Ferno CCT PX Critical Care Troll...

Страница 2: ...r safety measures taken by the user should be within and under consideration of applicable regulations and local protocol Training on the proper use of this product must be provided before using this...

Страница 3: ...ing SX Cotsides________________17 18 5 15 SX Cotside Positioning for Patient Lateral Transfer__ 18 5 16 SX Cotside Ratcheting Positions________________ 19 5 17 Using SX Cotsides for Bariatric Transfer...

Страница 4: ...se refer to service manual for instructions Trapping hazards are unavoidable in any design of Trolley due to its folding action Operate with care Safety Information WARNING Do NOT exceed the safe work...

Страница 5: ...ged labels immediately New labels are available from your distributor 1 7 Symbol Glossary The symbols defined below are used on the trolley and or in this user manual Ferno uses symbols recognised by...

Страница 6: ...2 Training Operator trainees need to Have followed a training plan Read and understand the instructions in this manual Practice using the fully equipped trolley with weight and under simulated condit...

Страница 7: ...lack Orange webbing Cross strap patient harness straps quick and easy to deploy Lateral folding cotsides Easy to operate lift assisted backrest and leg position platform Trendelenburg Shock Knee Conto...

Страница 8: ...ole and then tighten the pole clamp fully Once tightened recheck that the device is indeed correctly attached A number of specialist brackets for a range of medical devices are available from Ferno fo...

Страница 9: ...inciple as a belt buckle It is important that the straps are fully tightened at all times when a cylinder is located in the holder To remove the cylinder loosen the holder straps and then remove the c...

Страница 10: ...Oxylog 3000 ventilator can be attached to the trolley To attach simply push the unit into the mounting bracket until it locks in place To release pull down the metal flap under the unit and then pull...

Страница 11: ...avoid it lifting up out of place Attaching the IV pole to trolley side rail For more information on the 3 stage IV Pole please also refer to the 3 stage IV pole User Manual reference 00271 AU V01 whic...

Страница 12: ...y When PacRac is not in use the poles can be left on the trolley or can be removed according to preference The PacRac equipment platform is attached to the trolley by sliding it onto the fixing points...

Страница 13: ...orm allowing devices to be viewed at all times when moving patients instead of mounting these devices underneath the trolley The various medical device mounts can be supplied either with a swivel moun...

Страница 14: ...the desired angle by squeezing the lever and pushing down on the leg platform The end section of the leg platform can be positioned at any angle to achieve the required leg position from horizontal t...

Страница 15: ...and push down until it locks in place Make sure it is correctly locked in place by trying to pull the headrest back out The red slide button located at the base of the headrest on the rear of the bac...

Страница 16: ...l lock push brake pedal back up until it is horizontal again Wheel brake engaged pedal points down Wheel brake disengaged pedal raised 5 11 Directional Wheel Lock Option Directional wheel lock option...

Страница 17: ...tient sizes and which also ratchet to near horizontal giving an extended patient surface for larger bariatric patients with even larger physical body mass The SX cotsides have 3 ratchet positions plus...

Страница 18: ...locked in place Moving SX cotside whilst depressing lever Bring SX cotside past horizontal Release lever and move SX cotside back up to required position 6 OPERATING THE TROLLEY Operating the Trolley...

Страница 19: ...ical position 1 10o SX cotsides ratchet position 2 27 5o 5 16 SX Cotside Ratcheting Positions SX cotsides ratchet position 4 62 5o SX cotsides in near horizontal position 80o SX cotsides folded fully...

Страница 20: ...point of SX Surface Xtenders 6 OPERATING THE TROLLEY 5 18 Positioning of SX Cotsides The Surface Xtender cotsides normal positioning on CCT PX is such that the head end side of the cotside is in line...

Страница 21: ...e All equipment and articles inside the back of a vehicle must be safely secured on an appropriate crash tested bracket or in a cupboard to avoid them becoming a projectile in the event of a vehicle c...

Страница 22: ...ed to side Pull towards end of trolley to unlock then pull towards you to unfold Unfold fully until clicks in place Side handles ready for use 5 24 Manoeuvring Handles Option Available as an option ar...

Страница 23: ...des are placed in the fully down position on both sides of the stretcher SX cotsides moved to fully down position 5 25 Attaching Split Scoop Fixing Kit cont At the foot end of the trolley place one of...

Страница 24: ...nd of the trolley and the head end of the Scoop at the foot end of the trolley as per the below picture This ensures maximum usable extended patient surface Lock the bracket in place by pressing down...

Страница 25: ...end bracket Connect Scoop halves at head end with extension fixing pole Attaching 2 halves of Scoop together using extension fixing pole Operating the Trolley Features 5 25 Attaching Split Scoop Fixin...

Страница 26: ...ice lever in full up position locked IMPORTANT BEFORE USING Re check all 4 locking levers on the fixing brackets to make sure they are fully down and that the fixing arms are securely locked to the tr...

Страница 27: ...applied to the trolley 2 Place the trolley beside the patient and undo the patient straps moving them away from the area the patient will be lying 3 Ensure that the backrest is in the required positio...

Страница 28: ...r side of the trolley Repeat the same for the other Crossover strap Repeat the same for the other Crossover strap and then adjust the length of straps using the adjust buckles so the Cross Strap harne...

Страница 29: ...when the straps are adjusted to shortest length If the straps have been adjusted longer they can be clipped into the corresponding buckle on that side to avoid hanging down to the floor Adjustable ext...

Страница 30: ...hind the backrest 7 PREPARING TO TRANSPORT A PATIENT Preparing to Transport a Patient 7 8 Attaching a Child Harness PediMate Neomate Unlike traditional tubing framed trolleys such as Pegasus the lower...

Страница 31: ...be attached to CCT PX by first attaching one of these fixing tangs to each of the blue fixing straps supplied with the ACR harness as per below photos The ACR blue fixing strap is looped through and a...

Страница 32: ...stud at base of backrest either side of the trolley 7 11 Using BabyPod on CCT PX Please refer to the BabyPod user manual for the procedure for attaching and using BabyPod on a trolley CCT PX can be us...

Страница 33: ...sferring on and off a vehicle via a tail lift or ramp 1 Transporting a patient on the trolley should always be carried out with the cotsides raised and the Cross Strap harness and leg strap fastened a...

Страница 34: ...ARING TO MOVE THE TROLLEY Preparing to Move the Trolley Head end lock spike fixing U bracket Foot end lock fixing bracket 9 1 Locking Devices 2 Part Lock Locking devices available for use with CCT PX...

Страница 35: ...ontact time Ferno recommends you inspect the trolley for damage as you clean and disinfect it See note regarding disinfectants to use 1 Remove the mattress patient harness straps and any accessories 2...

Страница 36: ...0 3182 Split Scoop Extreme Bariatric extended patient surface kit 009008250 Inflatable sidepads kit for use with Split Scoop kit 23271000802 Trolley dust cover elasticated AB material 836353200 AB Pat...

Страница 37: ...and legs straps all present and properly installed Are the patient straps mattress and harness in good condition with no cuts or frayed edges Are patient strap buckles free of visible damage and do t...

Страница 38: ...FERNO MAN 1121 226 R3 UK 38...

Страница 39: ...12 months Below is a service schedule sheet where servicing dates and activities can be recorded Ferno UK offers a range of cost effective servicing programmes helping to keep your equipment in good c...

Страница 40: ...FERNO MAN 1121 226 R3 UK 40...

Отзывы: