- 86 -
INF-TN1FSH-E
4.4.20.2. Pulse Doppler
Description
The data required for pulse Doppler measurement can be set as follows.
[Outline drawing of signal processing]
Transmissio
n start
(TXS)
Reference count
Echo
wave
(SIG)
Demodulated
wave
(ADN
/
ADO)
Signal wait time
Window
(RWIN)
Repetition cycle
Repetition frequency = 1/repetition cycle
Number of channels
Sampling
frequency
Measurement window
A. Send signal
Echo wave
Within 3.2 Vpp
Transmission
pulse
Magnified view of
A Transmission signal
Trans-
mission
voltage
Transmission frequency
Note: Make the setting, following the description in “6.3. Checking received waveform.”
CAUTION
•
This parameter is intended for our service personnel.
•
Do not change the setting, since the parameter affects the flow rate measurement. If the setting is
changed, measurement may be disabled.
•
Make the setting when the factory-set value poses problems in flow rate measurement. If no problem
arises with the factory-set value, the setting is not necessary.
Enter data for each item (see the following table) according to the display.
Item
Input method
Function and range or menu
Setting of wedge sound velocity of the sensor (Factory-set value: AUTO)
Wedge sound
velocity
Selection
•
AUTO
•
MANUAL
Numeric
value
•
WEDGE S.V.: 1000m/s to 3700m/s
Setting of pipe sound velocity (Factory-set value: AUTO)
Selection
•
AUTO
•
MANUAL
Pipe sound
velocity
Numeric value
•
PIPE S.V.: 1000m/s to 3700m/s