- 50 -
INF-TN1FSH-E
#
Select “LINING MATERIAL”
LINING MATERIAL
NO LINING
ENTER
"
or
#
ENTER
Enter select/enter mode, select “TAR
EPOXY,” and press ENTER.
LINING MATERIAL
TAR EPOXY
#
Select “LINING THICKNESS”
LINING THICKNESS
0.01 mm
ENTER 1 . 2 5 ENTER
Enter numeric value enter mode, enter
“1.25” using ten keys, and press ENTER.
LINING THICKNESS
1.25 mm
#
Select “KIND OF FLUID”
KIND OF FLUID
WATER
ENTER
"
or
#
ENTER
Enter select/enter mode, select “OTHERS,”
and press ENTER.
KIND OF FLUID
OTHERS
#
Select “FLUID S.V.”
FLUID S.V.
1440 m/s
ENTER 1 3 8 8 ENTER
Enter numeric value enter mode, enter
“1388” using ten keys, and press ENTER.
FLUID S.V.
1388 m/s
#
Select “VISCOSITY”
VISCOSITY
1.0038 E
−
6
m
2
/s
ENTER 1 . 1 2 9 ENTER
Enter numeric value enter mode, enter
“1.129” using ten keys, and press ENTER.
VISCOSITY
1.1290 E
−
6
m
2
/s
ESC
Display SENSOR SPACING.
SENSOR SPACING
39.16 mm
ESC
Display the measurement, reflecting the
setting.
(Measurement display screen)