INF-TN1FSH-E
- 37 -
3.6.2. Total pulse output setting
Description
Make the setting to perform pulse output of the total measurement value (flow rate).
The following is an example of operation. Refer to 4.4.9. and 4.4.11. for details of the setting.
Definition of total pulse to DO
(example)
When DO1 is output as total pulse output in forward direction
Key operation
Description
Display
FUNC STATUS
Display Status.
SELECT STATUS
DO.1
ENTER
"
or
#
Select “DO.1.”
SELECT STATUS
DO.1
▲
▼
ENTER
Display “OUTPUT DO.1”
OUTPUT DO.1
NOT USED
ENTER
"
or
#
ENTER
Enter select/enter mode, select “F: TOTAL
PULSE,” and press ENTER.
OUTPUT DO.1
F: TOTAL PULSE
"
or
#
Display “MODE DO.1”
MODE DO.1
REVERSE
ENTER
"
or
#
ENTER
Enter select/enter mode, select “NORMAL,”
and press ENTER.
MODE DO.1
NORMAL
ESC ESC
Display the measurement, reflecting the
setting.
(Measurement display screen)
Total setting (example)
When performing integration, holding output burnout with total pulse output set to 100 m
3
for 1 pulse, burnout timer set to 15 sec, and pulse width 1 set as 100msec.
Key operation
Description
Display
FUNC TOTAL
Display Total mode.
TOTAL MODE
TOTAL RUN
ENTER
"
or
#
ENTER
Enter select/enter mode, select “TOTAL
STOP,” and press ENTER. (Setting is
allowed.)
TOTAL MODE
TOTAL STOP
"
or
#
Select “TOTAL UNIT”
TOTAL UNIT
mL
ENTER
"
or
#
ENTER
Enter select/enter mode, select “m
3
,” and
press ENTER.
TOTAL UNIT
m
3
"
or
#
Select “TOTAL RATE”
TOTAL RATE
10.000 m
3
ENTER 1 0 0 ENTER
Enter numeric value enter mode, enter “100”
using the ten key, and press ENTER.
TOTAL RATE
100.000 m
3
"
or
#
Select “OUTPUT BURNOUT”
OUTPUT BURNOUT
NOT USED
ENTER
"
or
#
ENTER
Enter select/enter mode, select “HOLD,”
and press ENTER.
OUTPUT BURNOUT
HOLD
"
or
#
Select “BURNOUT TIMER”
BURNOUT TIMER
10 sec
ENTER 1 5 ENTER
Enter numeric value enter mode, enter “15”
using ten keys, and press ENTER.
BURNOUT TIMER
15 sec
"
or
#
Select “PULSE WIDTH 1”
PULSE WIDTH 1
50 msec