ST100 Series Flow Meter
OPERATION
Fluid Components International LLC
31
External Control Group Switching (EGS)
This mode automatically changes the ST100 calibration group based on a 4-20 mA output current from another device fed into the ST100
Auxiliary Input port.
Start the application. Click
USB Connect
on the home screen. Select the
Configuration
branch from the menu tree on the window's left side.
Select the
Extended Op. Mode
tab. Click
Toggle System Mode
. Enter user level password
2772
at the prompt. Verify that instrument is in
the service state with
Setup
showing in
System Mode
field. Click
External Control Group Switching (EGS)
radio button.
External Control Group Switching (EGS) Setup
Select the
Group Switch Setup
tab. In the window’s
External Control Group Switching Setup
field, specify a 4-20 mA input value in the text
box and and its calibration group assignment via the associated
Group
drop-down list (define up to 5 points starting with the lowest value at
the bottom). These entries determine which calibration group is active as the auxiliary input current varies. The EGS setup field's
Reset
button
provides a quick way to clear out all programming entries (mA value/calibration group assignment).
Click
Send to Device
to send the EGS programming to the instrument. Return to normal operation by selecting the
Extended Op. Mode
tab
and clicking the
Toggle System Mode
button (observe that
System Mode
field shows
Running
).
The Process Data screen on the ST100 configuration application will now display “EXTENDED OP MODE EGS” in the upper left corner. The
HMI display also shows
EGS
on the Group Number line after the HMI is refreshed. Refresh the HMI display by cycling the power to the unit
or covering the Hot Key for 3 seconds.
Example Process Data Screen with EGS Mode Shown
Group Switching Criteria,
4-20 mA Input Current
Group Switching Based on
Input Current
Содержание ST100 series
Страница 12: ...GENERAL ST100 Series Flow Meter 6 Fluid Components International LLC Intentionally Left Blank ...
Страница 38: ...OPERATION ST100 Series Flow Meter 32 Fluid Components International LLC Intentionally Left Blank ...
Страница 40: ...MAINTENANCE ST100 Series Flow Meter 34 Fluid Components International LLC Intentionally Left Blank ...
Страница 54: ...TROUBLESHOOTING ST100 Series Flow Meter 48 Fluid Components International LLC Intentionally Left Blank ...
Страница 80: ...APPENDIX C ADDITIONAL INFORMATION ST100 Series Flow Meter 74 Fluid Components International LLC ...
Страница 81: ...ST100 Series Flow Meter APPENDIX C ADDITIONAL INFORMATION Fluid Components International LLC 75 ...
Страница 82: ...APPENDIX C ADDITIONAL INFORMATION ST100 Series Flow Meter 76 Fluid Components International LLC ...
Страница 92: ...APPENDIX D GLOSSARY ST100 Series Flow Meter 86 Fluid Components International LLC Intentionally Left Blank ...
Страница 94: ...APPENDIX E APPROVALS ST100 Series Flow Meter 88 Fluid Components International LLC Intentionally Left Blank ...