OPERATION
ST100 Series Flow Meter
30
Fluid Components International LLC
Select the
Group Switch Setup
tab. In the window’s
Auto FE Calibration Group Switching Setup
field, specify the process data type via the
Pro-
cess Data
drop-down list (choose from
Flow
,
Temperature,
or
Pressure
). Observe that the corresponding engineering units are shown below the
Process Data
selection. The application then automatically defines the range over which group switching occurs by populating the bottom and top
process data value text boxes with the calibrated MIN and MAX values respectively. Enter one or more process data values as required for the
application. Assign an FE calibration group to a process data value entry by selecting its number from the
FE Cal. Group
drop-down list. A process
data entry without a calibration group assignment assumes the last switched calibration group. The FCS setup field's
Reset
button provides a
quick way to set process data to
Flow
(with calibrated MIN and MAX values entered into the bottom and top process data value text boxes).
Click
Send to Device
to send the FCS programming to the instrument. Return to normal operation by selecting the
Extended Op. Mode
tab
and clicking
Toggle System Mode
(observe that
System Mode
field shows
Running
).
The Process Data screen on the ST100 configuration application will now display “EXTENDED OP MODE FCS” in the upper left corner. The
HMI display also shows
FCS
on the Group Number line after the HMI is refreshed. Refresh the HMI display by cycling the power to the unit or
covering the Hot Key for 3 seconds.
Auto FE Calibration Group Switching (FCS)
This mode automatically changes the ST100 calibration group based on specific process data values of flow rate, temperature, or pressure (in FCI Units).
Start the application. Click
USB Connect
on the home screen. Select the
Configuration
branch from the menu tree on the window's left side. Select
the
Extended Op. Mode
tab. Click
Toggle System Mode
. Enter user level password
2772
at the prompt. Verify that instrument is in the service
state with
Setup
showing in
System Mode
field. Click
Auto FE Calibration Group Switching (FCS)
radio button.
Auto FE Calibration Group Switching (FCS) Setup
Example Process Data Screen with FCS Mode Shown
Select Flow, Temperature,
or Pressure
Group Switching Criteria,
Process Value
Group Switching Based on
Process Value
Содержание ST100 series
Страница 12: ...GENERAL ST100 Series Flow Meter 6 Fluid Components International LLC Intentionally Left Blank ...
Страница 38: ...OPERATION ST100 Series Flow Meter 32 Fluid Components International LLC Intentionally Left Blank ...
Страница 40: ...MAINTENANCE ST100 Series Flow Meter 34 Fluid Components International LLC Intentionally Left Blank ...
Страница 54: ...TROUBLESHOOTING ST100 Series Flow Meter 48 Fluid Components International LLC Intentionally Left Blank ...
Страница 80: ...APPENDIX C ADDITIONAL INFORMATION ST100 Series Flow Meter 74 Fluid Components International LLC ...
Страница 81: ...ST100 Series Flow Meter APPENDIX C ADDITIONAL INFORMATION Fluid Components International LLC 75 ...
Страница 82: ...APPENDIX C ADDITIONAL INFORMATION ST100 Series Flow Meter 76 Fluid Components International LLC ...
Страница 92: ...APPENDIX D GLOSSARY ST100 Series Flow Meter 86 Fluid Components International LLC Intentionally Left Blank ...
Страница 94: ...APPENDIX E APPROVALS ST100 Series Flow Meter 88 Fluid Components International LLC Intentionally Left Blank ...