![FCI ST100 series Скачать руководство пользователя страница 34](http://html.mh-extra.com/html/fci/st100-series/st100-series_installation-operation-and-maintanance-manual_547358034.webp)
OPERATION
ST100 Series Flow Meter
28
Fluid Components International LLC
External Input Flow Adjust (EIA)
This mode corrects the flow rate of an ST100 flow meter when an external 4-20 mA current is fed into the Auxiliary Input port. Four polyno-
mial factors are used to determine the correction applied to the flow rate and corresponding 4-20 mA output. The equation that defines this
correction is as follows:
Corrected Flow Rate = (Starting Flow Rate x Factor1) + (Factor2 x 4-20 mA Input) + (Factor3 x 4-20 mA Input
2
) + (Factor4 x 4-20 mA Input
3
)
Make all necessary connections as shown in the wiring diagram in Figure B-19.
Start the application. Click
USB Connect
on the home screen. Select the
Configuration
branch from the menu tree on the window's left side. Se-
lect the
Extended Op. Mode
tab. Click
Toggle System Mode
. Enter user level password
2772
at the prompt. Verify instrument is in the service
state with
Setup
showing in
System Mode
field.
Click
External Input Flow Adjust (EIA)
radio button. In the window’s
Ext. Input Flow Adjust
Setup
field enter the four polynomial factors in the text box for
Factor1
,
Factor2
,
Factor3,
and
Factor4
.
At window bottom click
Set System Mode to ‘Running’ when Send
check box. Click
Send to Device
near check box to send the EIA program-
ming to the instrument. Observe return to normal operation with
Running
showing in
System Mode
field.
The Process Data screen on the ST100 configuration application will now display “EXTENDED OP MODE EIA” in the upper left corner. The
HMI display also shows
EIA
on the Group Number line after the HMI is refreshed. Refresh the HMI display by cycling the power to the unit or
covering the Hot Key for 3 seconds.
External 4-20 mA Correction Input Screen
Example Process Data Screen with EIA Mode Shown
EIA Select
Enter the Factors as needed
Select (Checked)
Содержание ST100 series
Страница 12: ...GENERAL ST100 Series Flow Meter 6 Fluid Components International LLC Intentionally Left Blank ...
Страница 38: ...OPERATION ST100 Series Flow Meter 32 Fluid Components International LLC Intentionally Left Blank ...
Страница 40: ...MAINTENANCE ST100 Series Flow Meter 34 Fluid Components International LLC Intentionally Left Blank ...
Страница 54: ...TROUBLESHOOTING ST100 Series Flow Meter 48 Fluid Components International LLC Intentionally Left Blank ...
Страница 80: ...APPENDIX C ADDITIONAL INFORMATION ST100 Series Flow Meter 74 Fluid Components International LLC ...
Страница 81: ...ST100 Series Flow Meter APPENDIX C ADDITIONAL INFORMATION Fluid Components International LLC 75 ...
Страница 82: ...APPENDIX C ADDITIONAL INFORMATION ST100 Series Flow Meter 76 Fluid Components International LLC ...
Страница 92: ...APPENDIX D GLOSSARY ST100 Series Flow Meter 86 Fluid Components International LLC Intentionally Left Blank ...
Страница 94: ...APPENDIX E APPROVALS ST100 Series Flow Meter 88 Fluid Components International LLC Intentionally Left Blank ...