9 TROUBLESHOOTING - Probable causes of malfunctions
PROBLEM
CAUSE
REMEDY
Isolator does not work
–
the electricity supply has
been cut off at the mains
–
Electronic board out of order
–
Blown fuse(s)
–
check the voltage input to
the isolator
–
Replace PCB
–
Replace fuses
Alarm: “HEPA filters check”
Main HEPA filter clogged
Replace HEPA filter
Alarm: “Min. Pressure alarm”
The suction motor-fan does not
work properly
Check the gloves integrity
Check F6 fuse on the power
card
Check that the suction duct is
not clogged
Check the suction motor-fan
HEPA filters are clogged.
Replace HEPA filters.
Alarm: “Minimum LAF alarm”.
The main motor-fan does not
work
Check the terminal voltage of
the power card of the main
motor-fan
Check F7 fuse on the power
card
Replace the power card
Replace the microprocessor
card
Alarm: “Black-out ”
Blackout
Check the feeding cable, the
connection plug/socket, the
power supply line
Press “Mute” [15] key to silence
the alarm
Alarm “Sensors failure”
Failure of sensor XX
Replace XX sensor.
Alarm: “No encoder input LAF”
No signal from LAF flow sensor
Replace the LAF sensor
Содержание ASEPTIC
Страница 45: ...a D GROUP Company 00254EN Rev 01 11 2012 45 13 IDIAGRAM FOR MAINTENANCE OPERATIONS...
Страница 46: ...a D GROUP Company 00254EN Rev 01 11 2012 46...
Страница 47: ...a D GROUP Company 00254EN Rev 01 11 2012 47 13 J FRONTAL DIAGRAM...
Страница 48: ...a D GROUP Company 00254EN Rev 01 11 2012 48 13 K SIDE DIAGRAM...
Страница 49: ...a D GROUP Company 00254EN Rev 01 11 2012 49 14 SUPPORT TABLE...
Страница 52: ...a D GROUP Company 00254EN Rev 01 11 2012 52 17 WIRING DIAGRAM ASEPTIC EQUIPPED WITH CLARUS L 2...
Страница 53: ...a D GROUP Company 00254EN Rev 01 11 2012 53...
Страница 54: ...a D GROUP Company 00254EN Rev 01 11 2012 54...
Страница 55: ...a D GROUP Company 00254EN Rev 01 11 2012 55...
Страница 56: ...a D GROUP Company 00254EN Rev 01 11 2012 56...
Страница 57: ...a D GROUP Company 00254EN Rev 01 11 2012 57...
Страница 58: ...a D GROUP Company 00254EN Rev 01 11 2012 58...
Страница 59: ...a D GROUP Company 00254EN Rev 01 11 2012 59...
Страница 60: ...a D GROUP Company 00254EN Rev 01 11 2012 60 18 WIRING DIAGRAM ASEPTIC EQUIPPED WITH STERIS VHP...
Страница 61: ...a D GROUP Company 00254EN Rev 01 11 2012 61...
Страница 62: ...a D GROUP Company 00254EN Rev 01 11 2012 62...
Страница 63: ...a D GROUP Company 00254EN Rev 01 11 2012 63...
Страница 64: ...a D GROUP Company 00254EN Rev 01 11 2012 64...
Страница 65: ...a D GROUP Company 00254EN Rev 01 11 2012 65...
Страница 66: ...a D GROUP Company 00254EN Rev 01 11 2012 66...
Страница 67: ...a D GROUP Company 00254EN Rev 01 11 2012 67...