Chapter 6: Maintenance and Troubleshooting
compensation. It simply indicates that the data collected was within
specification.
Mounting
Most accuracy and repeatability problems can be attributed to improper
mounting of the Gage. These issues are not always obvious. An apparently stable
mount can actually be moving more than the accuracy of the device, directly
impacting results and causing system performance degradation.
If using a magnetic mount, check that mating surfaces are
clean and free of debris.
N
OTE
: Small metal shavings can cause the Gage to rock back
and forth. The magnet will naturally attract these shavings.
08M47E00_FAROGage.book Page 47 Wednesday, June 8, 2011 4:12 PM
Содержание Gage
Страница 1: ...FARO Gage Manual MAY 2011 08M47E00_FAROGage book Page 1 Wednesday June 8 2011 4 12 PM...
Страница 2: ...08M47E00_FAROGage book Page 2 Wednesday June 8 2011 4 12 PM...
Страница 4: ...08M47E00_FAROGage book Page 4 Wednesday June 8 2011 4 12 PM...
Страница 8: ...08M47E00_FAROGage book Page iv Wednesday June 8 2011 4 12 PM...
Страница 32: ...08M47E00_FAROGage book Page 24 Wednesday June 8 2011 4 12 PM...
Страница 38: ...08M47E00_FAROGage book Page 30 Wednesday June 8 2011 4 12 PM...
Страница 60: ...08M47E00_FAROGage book Page 52 Wednesday June 8 2011 4 12 PM...
Страница 74: ...08M47E00_FAROGage book Page 66 Wednesday June 8 2011 4 12 PM...
Страница 78: ...08M47E00_FAROGage book Page 70 Wednesday June 8 2011 4 12 PM...
Страница 94: ...08M47E00_FAROGage book Page 6 Wednesday June 8 2011 4 12 PM...
Страница 102: ...08M47E00_FAROGage book Page 4 Wednesday June 8 2011 4 12 PM...
Страница 103: ...08M47E00_FAROGage book Page 1 Wednesday June 8 2011 4 12 PM...