Chapter 6: Maintenance and Troubleshooting
• Why does it take longer for my measuring software to open when I use the
FARO Wireless connection?
The Gage driver requires more time to initialize when using a FARO
Wireless connection and this translates into slightly increased load times
for the measuring software.
• Why does it take longer for the C
OMPENSATION
dialog box to go away when I
am using a FARO Wireless connection.
Probe data saves in the Gage and requires more time for this information
to transfer when using the FARO Wireless connection.
• What happens if I go out of range and lose my connection? What happens if I
turn off the Gage while it is connected using the FARO Wireless connection?
Once the Gage driver detects that the Gage is no longer connected, after
about 10 seconds, a message appears asking if you want to restore the
FARO Wireless connection. If you click No, the connection stops. If
you click Yes, attempts are made to restore the connection.
• Why do I continue to see the message asking if I want to continue to use the
FARO Wireless connection?
Either the Gage has lost power, or it is out of the 30 feet or 10 meters
wireless range. After correcting the problem, click Yes to restore the
connection.
FARO Gage Performance Verification Checklist
If you are experiencing accuracy or repeatability problems with your Gage or
ScanArm system, this section details proper setup techniques to help you identify
the potential sources of error.
It is important to clearly understand all the “do's and don'ts” in order to get the
most out of every Gage and ScanArm system.
Issues That Will Degrade Accuracy
Mounting
1
Deflection in mounting.
• Do not mount to any type of wooden surface.
• Do not mount to optical tables such as a “Vibraplane.” The steel sheet on the
top of these tables is thin and will deflect.
2
Folding tripods without using the stabilizing struts are unstable.
08M47E00_FAROGage.book Page 44 Wednesday, June 8, 2011 4:12 PM
Содержание Gage
Страница 1: ...FARO Gage Manual MAY 2011 08M47E00_FAROGage book Page 1 Wednesday June 8 2011 4 12 PM...
Страница 2: ...08M47E00_FAROGage book Page 2 Wednesday June 8 2011 4 12 PM...
Страница 4: ...08M47E00_FAROGage book Page 4 Wednesday June 8 2011 4 12 PM...
Страница 8: ...08M47E00_FAROGage book Page iv Wednesday June 8 2011 4 12 PM...
Страница 32: ...08M47E00_FAROGage book Page 24 Wednesday June 8 2011 4 12 PM...
Страница 38: ...08M47E00_FAROGage book Page 30 Wednesday June 8 2011 4 12 PM...
Страница 60: ...08M47E00_FAROGage book Page 52 Wednesday June 8 2011 4 12 PM...
Страница 74: ...08M47E00_FAROGage book Page 66 Wednesday June 8 2011 4 12 PM...
Страница 78: ...08M47E00_FAROGage book Page 70 Wednesday June 8 2011 4 12 PM...
Страница 94: ...08M47E00_FAROGage book Page 6 Wednesday June 8 2011 4 12 PM...
Страница 102: ...08M47E00_FAROGage book Page 4 Wednesday June 8 2011 4 12 PM...
Страница 103: ...08M47E00_FAROGage book Page 1 Wednesday June 8 2011 4 12 PM...