![Faro Gage Скачать руководство пользователя страница 40](http://html1.mh-extra.com/html/faro/gage/gage_manual_3510779040.webp)
FARO Gage
May 2011
32
Chapter 5: Accuracy
The Gage compensation becomes invalid (stop functioning within FARO's
published standards) if one or more of the following occurs:
• Bending of the transfer cases (long, tube sections).
• Damage to the encoder-bearing interface joints (where tubes, bearings and
encoders are mated).
• Failure of the encoder-electronics system (changes motion into electrical data).
• Excessive impact with another object or an end-stop. The impact may cause
the transfer cases to bend, the joint housings to deform, or the encoder-bearing
interface pin system to loosen.
Any of these problems require FARO’s Customer Service Department to
disassemble the Gage, repair the necessary item(s), and perform a factory
compensation. A factory compensation is necessary since the physical nature of
the Gage has been changed.
Gage Calibration
The Gage software has commands that allow you to perform tests that certify the
accuracy and repeatability of your Gage. The calibration commands allow you to
test the Gage using a holes, a ball bar, or a step gauge.
The values measured in all of these calibrations include 100% of the measured
points. We will always refer to this as the “Max. minus Min. over Two,” or the
Maximum measurement subtracting from the Minimum measurement divided by
two (Max.-Min.)/2.
If any test fails, print the results and repeat the calibration two more times. Print
all of the results and contact FARO’s Customer Service by Phone, Fax or E-Mail.
See “Technical Support” on page 67.
Repeatability
Perform the single point repeatability test to check the probe and repeatability
using a hole or a cone. You must use a firmly secured object with a hole at least
1mm smaller than the ball probe. If a cone hole is used, extreme care must be
taken to ensure that the probe is seated at each “measurement.”
• Use a calibrated ball probe. See
“Installing Probes” on page 11
and
.
• Run the Single Point Articulation Test (SPAT).
See “Single Point Articulation
08M47E00_FAROGage.book Page 32 Wednesday, June 8, 2011 4:12 PM
Содержание Gage
Страница 1: ...FARO Gage Manual MAY 2011 08M47E00_FAROGage book Page 1 Wednesday June 8 2011 4 12 PM...
Страница 2: ...08M47E00_FAROGage book Page 2 Wednesday June 8 2011 4 12 PM...
Страница 4: ...08M47E00_FAROGage book Page 4 Wednesday June 8 2011 4 12 PM...
Страница 8: ...08M47E00_FAROGage book Page iv Wednesday June 8 2011 4 12 PM...
Страница 32: ...08M47E00_FAROGage book Page 24 Wednesday June 8 2011 4 12 PM...
Страница 38: ...08M47E00_FAROGage book Page 30 Wednesday June 8 2011 4 12 PM...
Страница 60: ...08M47E00_FAROGage book Page 52 Wednesday June 8 2011 4 12 PM...
Страница 74: ...08M47E00_FAROGage book Page 66 Wednesday June 8 2011 4 12 PM...
Страница 78: ...08M47E00_FAROGage book Page 70 Wednesday June 8 2011 4 12 PM...
Страница 94: ...08M47E00_FAROGage book Page 6 Wednesday June 8 2011 4 12 PM...
Страница 102: ...08M47E00_FAROGage book Page 4 Wednesday June 8 2011 4 12 PM...
Страница 103: ...08M47E00_FAROGage book Page 1 Wednesday June 8 2011 4 12 PM...