Model 520 HT UCA
D01157216
Revision A, October 2015
20
Under the Time column, the user can STOP TEST or select SHUTDOWN
parameters for the Auto Shutdown feature. The user can select between Time (hr)
or Compressive Strength and adjust the target value using the Shutdown Value
control. When the test reaches the Shutdown Value the test will be stopped and
cooling automatically turned on.
By selecting the VIEW LIVE CHART tab at the top of the screen, the user can
view real-time data in chart form. From this screen, the user may print the chart at
any time by pressing the PRINT button located at the bottom of the screen. (Please
note that a printer must be connected to the instrument before trying to print.)
Pressing the PRINT button displays the full TEST VIEWER screen where the user
may select specific printing options such as header/footer information before
printing to any installed printer.
A complete table of numerical data can also be viewed here by selecting the DATA
tab located on the left side of the screen in the Test Viewer.
Return to the test controls by selecting CLOSE. This will not have any effect on the
actual testing process and can be opened / closed at any time.
The TEST SETUP tab is simply a review of parameters entered during the TEST
SETUP and cannot be changed.
If there is some question as to whether the instrument is producing the desired
ultrasonic waveform, the user may press the PRINT WAVEFORM button