Model 520 HT UCA
D01157216
Revision A, October 2015
23
2.
Apply compressed air pressure of 60-180 psi to the cell.
3.
Insert the cell into the instrument and connect the top transducer and
thermocouple cables.
4.
In Instrument Setup, select CALIBRATE TRANSIT TIME from the touch
screen menu. When the calibration screen appears, press the CALIBRATE
button. If the waveform appears acceptable, press DONE to save the new
calibration values. In general, the peak amplitude should be 600 or greater
with water. If an acceptable waveform is not displayed, check the
transducers and cables and recalibrate.
Transit time calibration is different from older UCAs. The steel calibration bar is
no longer required. The unit is now calibrated with water. It is imperative that
there be no air in the water during calibration or while a test is in progress.
Incorrect or improper calibration values are the most likely reason for the UCA to
report incorrect strength values. The calibration should be checked whenever the
transducers are removed or if the compressive strength values become suspect.