65
Cleaning and disinfection for the
parts shown here
Parts
Remarks
1. Rinse with warm tap water*.
2. Check for encrustations and if
necessary, soak these in hot
water, then repeat Step 1.
3. Immerse components in the
disinfection solution (for
duration, check instructions for
use of the solution).
4. Rinse component again to
remove disinfectant residues.
5. Visual inspection concerning
disinfectant residues.
Leave to dry on clean cloth
(e.g. a flattened towel).
*
Alternatively, all components
shown on this page (except for the
tracheostoma mask) can also be
cleaned in a dishwasher.
Please clean them separately
(not together with dirty dishes) and
select a washing program with a
temperature of 55° Celsius or
higher.
Boiling the components with 100°
Celsius hot water is also possible.
In case of
reprocessing the
inhalation device
for re-use
(different
patients!), this
component must
be exchanged.
Risk of infection!
In case of
reprocessing the
inhalation device
for re-use (different
patients!), this
component must
be exchanged.
Risk of infection!
In case of
reprocessing the
inhalation device
for re-use (different
patients!), this
component must
be exchanged.
Risk of infection!
In case of
reprocessing the
inhalation device
for re-use (different
patients!), this
component must
be exchanged.
Risk of infection!