PQ3450-EU-EN-V1.2-05/12
6
Real time data logger saves data to SD memory card for download to
PC (data files open directly to spreadsheet)
Datalogger
Sampling rate: From 2 seconds to 7200 seconds
Data Output
RS232 port
Operating Temperature
0 to 50
o
C (32 to 122
o
F)
Operating Relative Humidity
80% Relative Humidity max.
Power Supply
Eight (8) ‘AA’ 1.5VDC batteries or AC 9V power adapter
Power Consumption
Meter: 270 mA DC; Clamp: 22 mA DC
Weight
968g (2.1 lbs.) (with batteries)
Dimensions
225 X 125 X 64 mm (8.86 X 4.92 X 2.52“)
2.2 Electrical Specifications
(45 to 65Hz; 23±5°C)
Function
Range Resolution
Accuracy*
AC Voltage
10.0V to 600.0V
0.1V
± (0.5% range + 3 digits)
AC Current
0.2A to 6000A
0.001A to 1 A
± (0.5% range + 5 digits)*
Power Factor
0.00 to 1.00
0.01
± 0.04
Φ
Phase Angle
-180° to 180°
0.1°
± 1°
Frequency
45 to 65Hz
0.1Hz
0.1Hz
0.0 to 3.6MW
0.0 to 3.6MVA
Power
0.0 to 3.6MVAR
0.001k to 0.001M
± (1% range+8 digits)
0.0 to 3.6MWH
0.0 to 3.6MVAH
Watt Hour
0.0 to 3.6MVARH
0.001k to 0.001M
± (1% range+8 digits)
*Meter accuracy only. For system accuracy add the accuracy of the clamp used.
PFH (Power Factor Hours): Long Term Power Factor
For three phase/four wire and three phase/three wire configurations:
PF
∑
= (PF1 + PF2 + PF3) / 3
For single phase three wire configurations:
PF
∑
= (PF1 + PF2) / 2
Shop for Data Logging products online at:
www.
DataLoggerStore
.ca
1.877.766.5412