PQ3450-EU-EN-V1.2-05/12
13
5.
Press SHIFT again and the “SHIFT 2” icon will appear on the upper right side of the screen
and the second half of the File Name, which the user can customize as needed, will be
highlighted. Use the arrow keys to select a number between 001 and 0050 inclusive. .
6.
File Name Examples:
a.
1P201001:
1P2 is one phase by two wires, 01 is the folder number, and 001 is the file number
b.
1P301001:
1P3 is one phase by three wires, 01 the folder number, and 001 the file number
c.
3P301001:
3P3 is three phases by three wires, 01 the folder number, and 001 the file number.
d.
3P401001:
3P4 is three phases by four wires, 01 the folder number, and 001 the file number.
7.
Press the SHIFT key again to continue with the Setup Mode editing.
File Name Screens
4.5.3 Set the Sampling Time (datalogging rate) for recording onto the SD Memory Card
1.
In the Setup Mode, use the up and down arrow keys to scroll to the SAMPLING TIME field.
2.
Press the SHIFT key and the symbol "SHIFT1" will switch ON.
3.
Use the arrow keys to adjust the sampling time; the range is 2 to 7200 seconds.
4.
Press the SHIFT key again to return to Setup Mode editing.
4.5.4 Delete a file on the SD Memory Card
1.
In the Setup Mode, scroll to the DELETE FILE field using the up and down arrow keys.
2.
Press and hold the SETUP/ENTER key for at least 2 seconds and the indicators “Y” and
“N” will appear next to the DELETE FILE field. Use the right and left arrow keys to highlight
“Y” for YES or “N” for NO and then momentarily press the SETUP/ENTER key.
3.
If “Y” is selected, the current file will be erased. If “N” is selected, the delete process will be
aborted. When a file is deleted the field next to DELETE FILE will show 100%. When a
delete is aborted, the field shows 0%.
Shop for Data Logging products online at:
www.
DataLoggerStore
.ca