PQ3350 V5.2 11/10
13
Power Quality of a 3 Phase 4 Wire (3P4W) System
a. Turn the power on. Press the POWER and the 1
Φ
3
Φ
buttons to select the 3P4W system. The
type of system will be displayed at the left bottom corner in the LCD.
b. Connect the four test leads to the voltage terminals V1, V2, V3 and the V
N
(Neutral) of the
system.
c. Connect the test leads to L1, L2, and L3 of the 3P4W system.
d. Connect the three current probes to the power analyzer input terminal I1, I2, and I3.
e. Clamp on to the L1, L2, and L3 of the 3P4W system. Make sure the current flows from the front
of the current probe to the back.
f.
All of the system parameters will be shown in the LCD
For a description of each parameter, refer to section XIV. NOMENCLATURE.
Power Quality of a 3 Phase 3 Wire (3P3W) System