Preparation for use
a.
Press SETUP button to enter setup screen. Press SETUP again to select the item for setting
(the selected item will be displayed in reverse video).
After selecting the item, press
▲
or
▼
buttons to set up its value.
b.
After finishing setting up, press EXIT button to leave setup mode.
1.
Select the data for download:
H means harmonics;
H in reverse video means HOLD screen data (if the data is the one you want, you can press
HOLD button to display the data, and press HOLD button again to exit);
P
means power data which can be a reference for downloading, the data in order is 0~84.
2. Shows the total logged data in the analyzer: max. 85 logging data.
3.
REC DATE:
shows the starting logging time of the 1
st
downloaded file.
4.
Hz:
set up the frequency (50, 60 or AUTO) of the system.
5.
PT:
set up PT value.
6.
CT:
set up CT value.
7.
SEC:
set up the interval seconds of logged data.
8.
CLAMP:
set up the clamps selected (100A, 1000A or 3000A).
9.
MD TIME:
set up the time of Maximum Demand (1~60 minutes).
10.
TRANS REF:
set up the transient voltage (which will be automatically changed in accordance with
PT).
11.
SDVP:
set up upper and lower limits % of transient voltage detection.
12.
YEAR:
Set up “year” of calendar clock.
13.
MONTH:
Set up “month” of calendar clock.
14.
DATE:
Set up “date” of calendar clock.
15.
HOUR:
Set up “hour” of calendar clock.
16.
MINUTE:
Set up “minute” of calendar clock.
17.
SECOND:
Second can be displayed only (it can not be adjusted).
1 2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17