PQ3350 V5.2 11/10
40
in AUTO mode
Model PQ3110 or PQ3120
Range
Resolution
Accuracy of Reading
45 – 65 Hz
0.1Hz
0.1Hz
Frequency
of ACV
(RMS value > 10V)
or ACA
(RMS value > 30A)
Model PQ3220/PQ3210
Range Resolution
Accuracy
45 – 65 Hz
0.1 Hz
± 0.2Hz
Total Harmonic Distortion
(THD-F with respect to the fundamental frequency, min. value at 50 or 60 Hz is
voltage > AC 80V and current is: model PQ3110> 10% of the range;
model PQ3120> 20A; model PQ3220/PQ3210> 30A.
Calculation is done over 1 to 50
th
Harmonics. If the voltage or current is 0 at 50 or
60 Hz, all the percentage (%) display is 0).
Model PQ3110
Range Resolution
Accuracy
0.0 – 20.0 %
± 1%
20.0 – 100%
±3% of reading ± 5%
100 – 999.9%
0.1%
±10% of reading ±10%
Model PQ3120
Range Resolution
Accuracy
0.0 – 20%
± 2%
20 – 100%
± 6% of reading ± 1%
100 – 999.9 %
0.1%
± 10% of reading ± 1%
Model PQ3220/PQ3210
Range Resolution
Accuracy
0.0 – 20%
0.1%
± 2%
20 – 100%
0.1%
± 6% of reading ± 5%
100 – 999.9 %
0.1%
± 10% of reading ± 10%