AC AND DC CURRENT MEASUREMENTS.
1. Turn the rotary switch to the
AC/DC A
position.
2. Press the
HELP
key and then the “
F4
” key to display
procedures, cautions and connection diagrams
regarding current measurements
3. Maximum allowed current is 20 Amps. Use a current
clamp for larger currents.
4. Press F3 to change from DC to AC current.
5. Press F4 to change to dB.
RPM MEASUREMENTS.
1. Turn the rotary switch to the RPM position.
2. Press the
HELP
key and then the “
F4
” key to display
procedures, cautions and connection diagrams
regarding rpm measurements
3. Connect the optional inductive pickup as described.
4. Press F1 for 2 cycle engines and F2 for 4 cycle
engines..
SIG.OUT
1. Turn the rotary switch to the SIG.OUT position.
2. Press the
HELP
key and then the “
F4
” key to display
procedures, cautions and connection diagrams
regarding the SIG.OUT function.
3. Press F1 to increase the frequency and F2 to
decrease the frequency.
4. Press F3 to initiate the output. Press again to disable
the output.
AUX
The AUX selection allows direct display of
o
C,
o
F, %RH, psi or High current in the
correct units when the meter is used with an optional adaptor.
1. Turn the rotary switch to the AUX position.
2. Press the
HELP
key and then the
F4
key to display procedures, cautions and
connection diagrams regarding the AUX function.
3. Press F1 to F4 to select the units that match the adaptor being used
INSTALL
The Install function provides a means to set three default
Power-on conditions (dB reference, key beeper, auto-
shutoff) and also enable self-test.
1. Turn the rotary switch to the
INSTALL
position.
2. Press the F1 key to select one of the three features.
3. Press F2 to enable or disable the feature or to set the
dB reference level.
4. Press F3 to save the changes.
5. Press F4 to run the self test.