Frequency Measurements 

1. Connect and make the measurement required as described in the 

previous paragraphs. 

2. Press the Hz key to select the Frequency (Hz) function. 

3. Read the frequency on the display  

Notes on sensitivity: 

Input sensitivity varies automatically with function range selected. The 

mV function has the highest (300mV) and the 1000V range has the 
lowest (300V) sensitivity. It is recommended that the user first 
measure the signal voltage (or current) level before activating the Hz 
function to automatically set the most appropriate trigger level. You 
can also press the RANGE button momentarily to select another 
trigger level manually. If the Hz reading becomes unstable, select a 
lower sensitivity to avoid electrical noise. If the display is zero, select 
a higher sensitivity. 

Diode Test 

1. Insert the black lead into the negative 

COM

 jack and the red test lead 

into the positive jack 

2. Set the function switch to "    

CAP

" position. 

3. Press the SELECT key momentarily to select the diode function (if 

required). 

4. Touch  the  test probe tips to the diode or semiconductor junction you 

wish to test. Note the meter reading. 

5. Reverse the test lead polarity by reversing the red and black leads. 

Note this reading. 

6. The diode or junction can be evaluated as follows: 

A. If one reading displays a value and the other reading displays "OL", 

the diode is good. 

B. If both readings display "OL", the device is open. 
C. If both readings are very small or 0, the device is shorted. 

Temperature Measurements 

1. Insert the Temperature Probe into the 

COM

 and 

TEMP

 input jacks 

observing polarity. 

2. Set the function switch to the 

"TEMP"

 position. 

3. Use the SELECT key to toggle between 

o

C and 

o

F. 

4. Touch the Temperature Probe tip to the device under test. 
5. Wait 30 seconds for the measurement to stabilize then read the 

display.  

Содержание MultiLog ML720

Страница 1: ...User s Guide MultiLog Digital MultiMeter With built in Datalogger Model ML720...

Страница 2: ...normal use hazardous voltages may be present Double insulation Safety Precautions 1 Improper use of this meter can cause damage shock injury or death Read and understand this user s manual before oper...

Страница 3: ...Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in which measures are taken to limit the transient overvoltages to an appropriate low level Note Examples include protected...

Страница 4: ...1k Input Impedance 10M 16pF nominal 44pF nominal for 50mV 500mV ranges DC CURRENT RANGE Accuracy Burden Voltage 500 0 A 5000 A 0 15mV A 50 00mA 500 0mA 3 3mV mA 5 000A 10 00A 0 2 4d 0 03V A 10A conti...

Страница 5: ...03V A 10A continuous 20A for 30 second max with 5 minute cool down interval TEMPERATURE RANGE Accuracy 50 C TO 1000 C 0 3 3d 58 F TO 1832 F 0 3 6d RESISTANCE RANGE Accuracy 50 00 0 2 6d 500 0 0 1 3d...

Страница 6: ...forms Audible Continuity Measurement threshold Beeper will sound if measurement is below 20 Beeper will not sound if measurement is above 200 Beeper may or may not sound if measurement is between 20 a...

Страница 7: ...time this manual was written UL Certification Meters that bear the UL marking have been investigated by UL headquarter in the USA per UL standard UL3111 1 1st Ed to its highest CAT III rating and int...

Страница 8: ...Description 1 5000 count Liquid Crystal display 2 Function push buttons 3 Rotary function switch 4 10A input jack 5 V Hz Cap Temp input jack 6 COM input jack 7 mA uA current input jack 8 RS232 connect...

Страница 9: ...The MultiLog is equipped with a backlight for viewing the LCD display in poorly lit areas Data Hold The HOLD function freezes the reading on the display for later viewing Manual and Automatic Ranging...

Страница 10: ...AC DC Current Measurements CAUTION Do not make current measurements at 20A for longer than 30 seconds Exceeding 30 seconds may cause damage to the meter and or the test leads 1 Insert the black test l...

Страница 11: ...est lead resistance calibration When manually entering the 50 range using the RANGE key the automatic test lead resistance feature will prompt Shrt in the display the user to short the inputs for cali...

Страница 12: ...Insert the black lead into the negative COM jack and the red test lead into the positive jack 2 Set the function switch to CAP position 3 Press the SELECT key momentarily to select the diode function...

Страница 13: ...t the desired sample rate 3 Press the timer key to store the selected sample rate 4 Press and hold the Strt PAUS StoP key for 1 second until Strt appears in the display to start the data logging proce...

Страница 14: ...ll of the trends in the recorded data list The LCD will indicate MIN or MAX for each trend in the list Datalog Item Numbers Each logged reading has an item number 0 to n where n is the maximum number...

Страница 15: ...mpling Rate cannot be changed while the meter is datalogging When a Sampling Rate of 20 seconds or longer is selected the meter enters a standby state between readings to conserve battery life While i...

Страница 16: ...open case Install only the same type of fuse or equivalent Cleaning and Storage Periodically wipe the case with a damp cloth and mild detergent do not use abrasives or solvents If the meter is not to...

Отзывы: