background image

MA220-EU-EN V1.5 7/09

3

Specifications

Function

Range

Accuracy (of
reading)

DC Current

40.00 ADC

0-20.00 ADC

± (2.5% + 6 digits)

20.00-40.00 ADC

± (3% + 6 digits)

400.0 ADC

0-300.0 ADC

± (2.5% + 6 digits)

300.0-400.0 ADC

± (3.5% + 6 digits)

AC Current

40.00 AAC

0-20.00 AAC

± (3% + 10 digits)

20.00-40.00 AAC

± (5% + 10 digits)

400.0 AAC

0-300.0 AAC

± (3% + 10 digits)

300.0-400.0 AAC

± (5% + 10 digits)

DC Voltage

400.0mV

± (0.8% + 3 digits)

4.000V

± (1.5% + 3 digits)

40.00V
400.0V
600V

± (2.0% + 3 digits)

AC Voltage

400.0mV

± (1% + 10 digits)

4.000V

± (2% + 5 digits)

40.00V
400.0V
600V

± (2% + 5 digits)

Resistance

400.0

± (1.0% + 4 digits)

4.000k

± (1.5% + 2 digits)

40.00k

400.0k

4.000 M

± (2.5% + 3 digits)

40.00M

± (3.5% + 5 digits)

Capacitance

40.00nF

± (5% + 30 digits)

400.0nF

± (3% + 5 digits)

4.000µF

± (3.5% + 5 digits)

40.00µF
100.0µF

± (5% + 5 digits)

Frequency

5.000Hz

± (1.5% + 5 digits)

50.00Hz

± (1.2% + 2 digits)
Sensitivity: 5~5KHz:
10Vrms min.
5KHz~150KHz:
40Vrms min.

500.0Hz
5.000KHz
50.00KHz
150.0KHz

Duty Cycle

0.5% to 99.0%

± (1.2% + 2 digits)

Pulse Width: 100µs-100ms, Frequency: 5Hz to 150KHz

Temperature

-50.0 to 400.0°C

-50.0 to -20.0°C

± 7°C

-20.0 to 400.0°C

± (3% + 5 °C)

400 to 1000°C

400 to 1000°C

-58.0 to 400.0°F

-58.0 to 0°F

± 14°F

0 to 400.0°F

± (2.5% + 6 digits)

400 to 1832°F

400 to 1832°F

± (3% + 7°F)

Содержание MA220

Страница 1: ...User s Guide 400A AC DC Clamp Meter Model MA220...

Страница 2: ...ltage to meter when resistance function is selected Set the function switch OFF when the meter is not in use Remove the battery if meter is to be stored for longer than 60 days WARNINGS Set function s...

Страница 3: ...000V 2 5 digits 40 00V 400 0V 600V 2 5 digits Resistance 400 0 1 0 4 digits 4 000k 1 5 2 digits 40 00k 400 0k 4 000 M 2 5 3 digits 40 00M 3 5 5 digits Capacitance 40 00nF 5 30 digits 400 0nF 3 5 digit...

Страница 4: ...per second nominal Input Impedance 7 8M V DC and V AC Operating Temperature 5 C to 40 C 41 F to 104 F Storage Temperature 20o C to 60o C 4o F to 140o F Operating Humidity Max 80 up to 31 C 87 F decre...

Страница 5: ...Mode select button 8 Range select button 9 Hz Duty Cycle button 10 COM input jack 11 V Temp jack 12 Battery cover rear AC AC alternating current DC DC direct currrent Minus sign AUTO AutoRange mode ZE...

Страница 6: ...positive V Temp jack 3 Select AC or DC with the MODE button 4 Connect the test leads to the circuit under test 5 Read the voltage on the display The display will indicate the proper decimal point and...

Страница 7: ...uty Cycle Measurements 1 Turn the rotary switch to the Volts Hz position 2 Insert the black test lead banana plug into the negative COM jack and the red test lead banana plug into the positive V jack...

Страница 8: ...the meter before opening the case Do not operate with open case Cleaning and Storage Periodically wipe the case with a damp cloth and mild detergent do not use abrasives or solvents If the meter is n...

Отзывы: