Diode Test

1. Turn the rotary switch to the

•))) CAP

position.

2. Insert the black test lead banana plug into the negative (COM) jack and the red test lead

banana plug into the positive (V



jack.

3. Push the mode button to indicate

on the display.

4. Touch the test probes to the diode under test. Typically for a normal diode, forward

voltage will indicate 0.4V to 0.7V. Reverse voltage will indicate “OL”. Shorted devices will
indicate near 0V and an open device will indicate “OL” in both polarities.

Capacitance Measurements

Warning

: To avoid electrical shock, disconnect power to the unit under test and

discharge all capacitors before taking any capacitance measurements. Remove the
batteries and unplug the line cords.

1.

Set the function switch to the

•))) CAP

position.

2.

Push the mode button to indicate

nF

on the display.

3.

Insert the black lead banana plug into the negative (COM) jack and
insert the red test lead banana plug into the positive (V

Temp) jack.

4.

Press the ZERO key to null the meter display.

5.

Touch the test probe tips to the capacitor you wish to check.

6.

Read the capacitance value on the display.

Frequency or % Duty Cycle Measurements

1.

Turn the rotary switch to the

Volts Hz %

position.

2.

Insert the black test lead banana plug into the negative (COM) jack and the red test lead
banana plug into the positive (V



jack.

3.

Select Hz or % with the

HZ/%

button.

4.

Touch the test probe tips to the circuit under test.

5.

Read the frequency on the display.

Temperature Measurements

1.

Turn the rotary switch to the

Temp

position.

2.

Insert the Temperature Probe into the negative (COM) and the (V

Temp

jacks, making

sure to observe correct polarity.

3.

Select °C or °F with the

MODE

button.

4.

Touch the temperature probe head to the part whose temperature you wish to measure.

Keep the probe touching the part under test until the reading stabilizes.

5.

Read the temperature on the display.

Warning

: To avoid electrical shock, be sure the thermocouple has been removed before

changing to another measurement function.

Auto/Manual Ranging

The meter turns on in Autoranging mode. Press the

RANGE

button to enter manual

ranging. Each press of the range button will step to the next range as indicated by the units
and decimal point location. Press and hold the

RANGE

button for two seconds to return to

Autoranging mode.

Note

: Manual ranging does not function in AC Current or Diode and Continuity check

functions. In Temperature function, it will change the resolution from 0.1° to 1°.

Содержание MA220

Страница 1: ...User s Guide 400A AC DC Clamp Meter Model MA220...

Страница 2: ...ltage to meter when resistance function is selected Set the function switch OFF when the meter is not in use Remove the battery if meter is to be stored for longer than 60 days WARNINGS Set function s...

Страница 3: ...000V 2 5 digits 40 00V 400 0V 600V 2 5 digits Resistance 400 0 1 0 4 digits 4 000k 1 5 2 digits 40 00k 400 0k 4 000 M 2 5 3 digits 40 00M 3 5 5 digits Capacitance 40 00nF 5 30 digits 400 0nF 3 5 digit...

Страница 4: ...per second nominal Input Impedance 7 8M V DC and V AC Operating Temperature 5 C to 40 C 41 F to 104 F Storage Temperature 20o C to 60o C 4o F to 140o F Operating Humidity Max 80 up to 31 C 87 F decre...

Страница 5: ...Mode select button 8 Range select button 9 Hz Duty Cycle button 10 COM input jack 11 V Temp jack 12 Battery cover rear AC AC alternating current DC DC direct currrent Minus sign AUTO AutoRange mode ZE...

Страница 6: ...positive V Temp jack 3 Select AC or DC with the MODE button 4 Connect the test leads to the circuit under test 5 Read the voltage on the display The display will indicate the proper decimal point and...

Страница 7: ...uty Cycle Measurements 1 Turn the rotary switch to the Volts Hz position 2 Insert the black test lead banana plug into the negative COM jack and the red test lead banana plug into the positive V jack...

Страница 8: ...the meter before opening the case Do not operate with open case Cleaning and Storage Periodically wipe the case with a damp cloth and mild detergent do not use abrasives or solvents If the meter is n...

Отзывы: