8.2 Power Up Failure
If the unit fails to POWER up or function properly, use the following checklist to
eliminate the most common causes of problems.
Check that:
1. The AC POWER cord is securely plugged into a functional AC wall plug.
2. The AC POWER cord is securely plugged into the AC inlet on the rear of the unit.
3. The main AC POWER switch is in the ON position.
4. Check that the ventilation openings on both the bottom and rear of the unit are not
blocked.
If the unit still does not power-up:
5. Check both main power fuses by first disconnecting the power cord. Then carefully
remove the fuse drawer assembly below the AC inlet on the rear of the unit. If the fuse(s)
is/ are open, replace with the same type (5A, 250 V, Fast acting).
8.3 Lamp Strike Failure
If the LED display lights and the fan starts, but the lamp will not turn on, check for
the following:
1.
The LED display indicates the “
bulb
” message and the system begins to beep. This
indicates that no lamp is detected. Check if the lamp has been installed correctly. Refer
to
section 5.1
– Installing the Lamp Module.
2.
The LED display indicates the “
bulb
” message after approximately 45 seconds and
the system will beep 3 times as an audible warning. This indicates that the lamp has
failed to strike. It may be a result of the lamp reaching end of life, or that the lamp
housing panel is not secured properly in place. Press the MODE button to clear the
audible alarm. Turn power off to the unit. Check that the lamp housing panel is secured
properly. Wait a few minutes and turn the power on to the unit. If it still does not strike,
replace the lamp.
3.
The LED display indicates the “
Cool
” message. This indicates the lamp is too hot to
strike. The lamp will automatically strike when it has cooled.