18
X-Cite
®
200DC User Guide
035-00446R rev. 4
6.1.6 To power down the unit after use, set the main power switch located on the front
panel
to “0”.
Note:
If the lamp is turned off, and an attempt is made to turn it back on before it
has fully cooled, the “
Cool
” message will appear on the display and the
lamp indicator will be un-lit. The lamp will automatically re-strike when the
lamp has cooled, the lamp indicator will illuminate (green).
6.2 Illuminating a Specimen
6.2.1 When the X-Cite
®
200DC is first powered on, the internal shutter will be in the
closed position as a safety precaution. If the shutter is not closed error code
SFI
will appear on the LED display.
6.2.2 To illuminate a specimen, press and release the
SHUTTER
button to open the
shutter (
OPEN
LED will turn on).
6.2.3 Use
“Iris Adjustment Dial”
for intensity adjustment to desired level.
6.2.4 Press the
SHUTTER
button to close the shutter. The
OPEN
LED will turn off.
6.3 Front Panel Button Functions
Button
Condition
Description
I/O
Press on press off
Turns unit on “I” Turns unit off “0”
SHUTTER
Press and release
Opens/closes internal shutter
6.4 LED Indicators
LED Indicator
Status
Description
LIGHT GUIDE
On
– green
The light guide is fully inserted
On
– red
The light guide is not fully inserted
LAMP
On - green
The lamp is on
Off
– no illumination The lamp is off
OPEN
On - green
The shutter is open
Off
– no illumination The shutter is closed
Содержание X-Cite 200DC