21
X-Cite
®
200DC User Guide
035-00446R rev. 4
5. To align the detection window with the visible light from the microscope objective,
use the microscope stage x-y controls, ensuring the light is centered within the
window.
6. Engage the appropriate objective, filter cube, intensity and other microscope
settings and press the
‘
(nm)’
button on the XR2100 to see the current
wavelength setting.
XP750 Objective Plane Power Meter Sensor (
with inverted microscopes);
1. Plug the XP750 multi-pin connector into the XR2100 turn the XR2100 to ON.
2. Place the XP750 onto the stage with the detection window facing down, and
approximately centered beneath the microscope objective lens.
The + icon
(etched inside a circle) should be facing up which marks the center of the
detection window.
3.
Secure the XP750 in place using the microscope’s stage clip and the supplied
suction-cup ensuring the XP750 is seated firmly and turn on the transmission
light source and using the microscope stage x-y controls, center the light on the +
icon.
4. Engage the appropriate objective, filter cube, intensity and other microscope
settings and press the
‘
(nm)’
button on the XR2100 to see the current
wavelength setting.
Note: