13.4.5 Exposure Settings: Disabling Exposure Timer
To disable the exposure timer,
de-select
the checkbox next t
o “Exposure timer
active
”. This will allow the shutter to remain in the open or closed position as per
the shutter control, without any set timing function.
13.4.6 Shutter Control
To activate the shutter control once the appropriate intensity and exposure
settings have been entered,
click on the “Open” button in the “Shutter Control”
section.
After the exposure cycle has been completed, the shutter will close and the
exposure time will automatically reset. The Unit Status indicator will display
Shutter “Close”. To close the shutter during an exposure cycle, click on the
“Close” button.
13.4.7 Alarm Function
The following conditions will cause the unit to indicate an alarm condition:
i) Lamp fail to strike
ii) Lamp access door not correctly installed
iii) Intelli-lamp sensor/ connection failure
iv) Shutter position failure
The alarm condition is indicated by a repeating audible beep from the X-Cite
®
120PC Q unit. The
Unit
Status
indicator on the GUI will display
Alarm
“On”.
To clear the alarm signal, click on the “Clear Alarm” button. The audible alarm on
the unit will cease and the
Unit Status
indicator will display
Alarm “Off”.
In addition to clearing the alarm signal on the GUI, the X-Cite
®
120PC Q unit
must have the front panel power switch reset to clear the alarm condition. The
fault which caused the original alarm condition must be corrected prior to
resuming normal operation.
Note:
During a timed exposure cycle, when the shutter is open, a red bullet
will appear next to “Exposure”. Unit Status indicator will display Shutter
“Open”.
Note:
The “Unit Status” indicators on the GUI may be beneficial in diagnosing
any potential causes relating to the alarm condition. Refer to Section 13
for additional information on troubleshooting.