![Excelitas Technologies X-Cite 120PC Q Series Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/excelitas-technologies/x-cite-120pc-q-series/x-cite-120pc-q-series_user-manual_2449452035.webp)
30
X-Cite
®
120PC Q User Guide
035-00328R rev. 4
6. " end / bulb"
The “end” and then “bulb” message appear in alternating sequence when power is first
turned on to the unit if the accumulated lamp hours have exceeded 4000 hours. This
indicates that the lamp has reached the end of safe life. If the lamp hours reach 4000
hours, safety measures will prevent the lamp from striking. The lamp must be replaced
at this point.
7.
"cool"
The "cool" message advises the user that the lamp must cool down before re-ignition. It
appears on the display when power is turned on to the unit and the lamp is too hot to
strike. Striking the lamp when it is too hot will reduce lamp life. The lamp will
automatically strike once it has reached optimum temperature.
8.
"LOC"
The LOC message appears for about 2 seconds when the
UP
/
DOWN
adjustments are
initially locked. It also appears if the up or down button is pressed while they are locked,
advising the user that these functions are locked out. See
Section 11.2
– Locking and
Unlocking the
UP
/
DOWN
Adjustment Buttons.
9.
"ULOC"
The ULOC message appears for about 2 seconds after unlocking the
UP
/
DOWN
adjustment buttons to advise the user they have been unlocked. See
Section 11.2
Locking and Unlocking the
UP
/
DOWN
Adjustment Buttons.
10. "SFI"
The “SFI” message appears if there is a shutter failure. Power down the unit, wait a
minute and turn power on to the unit. If the unit displays a shutter failure again, contact
your local X-Cite Service Centre to have your unit serviced.
1. LAMP
The
LAMP
indicator is the top of two green LED’s located to the left of the display. The
LAMP
LED illuminates when the lamp is on. The
LAMP
LED will flash when the lamp is
warming up.
2. OPEN
The
OPEN
indicator is the bottom of two green LED’s located to the left of the display.
The
OPEN
LED illuminates when the shutter is open.