13
X-Cite
®
120PC Q User Guide
035-00328R rev. 4
4.1.10 Replace the lamp access panel and tighten the fastening screw. Reference
.
5
Inserting and Removing the Light Guide
5.1.1 Make sure that power is off to the X-Cite
®
120PC Q unit.
5.1.2 Ensure that the protective end caps are removed from both the input and output
ends of the light guide prior to installation.
5.1.3 Insert the light guide into the light guide retainer located on the back panel of the
unit. Push the light guide in until it seats with a second positive "click" (up to the
“A” line printed on the grey ferrule). Reference Figure 9.
Figure 9 Light Guide Insertion
Warning
This unit is designed for bench top use only! Always ensure that the unit
is operated on a hard, stable surface. This will prevent obstruction of the
bottom chassis ventilation openings. Any obstruction of these openings
could result in a possible over-heating condition. Do not attempt to
remove or tamper with the rubber feet located on the bottom of the unit.
Note:
If the lamp module has been installed incorrectly the message
“bulb” will appear on the display and a continuous audible beep
will be heard after the 90 second warm-up cycle has been
completed.