User Guide
- 56
–
23.3 I/O Ports (including RS-232)
23.3.1
The S2000 PRO/ELITE has 5 different I/O ports; 4 located on the back of the
unit. The 1st port is a simple 2-wire, audio style jack that can be connected to a
foot pedal (supplied with each unit) or any other electro-mechanical triggering
device. This is an exposure trigger input used to start an exposure. The 2nd
port is a RS-232 port (9 pin) to be connected to a PC and is described in more
detail in section 19. The 3rd and 4th ports (15 pin DE style) are for PLC
interfacing. A 5thport is a RS-232 port located on the side of the front panel
which is used to connect to the R2000 Radiometer for calibration of the S2000
PRO/ELITE.
23.4 USB Interface
23.4.1
Blank
23.5 Environmental Conditions
Operating Conditions
Ambient Temperature:
15ºC to 40ºC
Altitude:
2000m max.
Atmospheric Pressure:
700 to 1060 hPa
Relative Humidity:
15% to 95% (non-condensing)
Installation Category:
II
Pollution Degree:
2
Transport and Storage Conditions
Temperature:
-40 to +70ºC
Relative Humidity:
10% to 100%
Atmospheric Pressure:
500 to 1060 hPa