3.  There are no bends, kinks, or other physical damage to the guide. Replace 

the light guide if there is any physical damage. 

4.  The lamp has been installed correctly. See section 6 

– Installing the Lamp 

Module. 

 

It may be necessary to replace the lamp or to replace the light guide. 
Contact your 

Excelitas Canada 

sales representative for information on 

purchasing a new lamp or light guide. 

 

22.1.5 

If the LCD display does not light: 

1.  If the fan is functional, POWER down the unit, wait approximately 20 

seconds then POWER it up again 

2.  If the problem persists, contact your local 

Excelitas Canada

 Service Centre. 

 

22.1.6 

If one or more fans do not work: 

1.  If the LED display is functional, POWER down the unit, wait approximately 

20 seconds then POWER it up again 

2.  If the problem persists, contact your local 

Excelitas Canada 

Service Centre. 

 
 

 

 

Содержание OmniCure S2000 Elite

Страница 1: ...I OmniCure S1500 PRO S2000 PRO ELITE USER S GUIDE Excelitas Canada Inc 2260 Argentia Road Mississauga ON L5N 6H7 Canada 1 905 821 2600 www excelitas com Printed in Canada Draft Rev 5...

Страница 2: ...y form by any means without the prior written consent of Excelitas Canada Inc Every effort has been made to insure information in this manual is accurate however information in this manual is subject...

Страница 3: ...nlocking the UP DOWN Adjustment Button 24 12 Timed Exposures 25 12 1 Adjusting the Exposure Time 25 12 2 Running a Timed Exposure 25 13 Interfacing with the S2000 PRO ELITE 26 13 1 Connection Options...

Страница 4: ...nmental Conditions 56 23 6 Noise and the OmniCure S2000 PRO ELITE 57 24 Regulatory Compliance 59 24 1 Product Safety and Electromagnetic Compatibility 59 24 2 FCC Part 15 Subpart B Class A Unintention...

Страница 5: ...Signal Circuit Configuration 27 Figure 8 Rear Panel Sample Connection 28 List of Tables Table 1 P1 Pin Out 29 Table 2 P3 Pin Out 30 Table 3 RS 232 Phono Connector Pin Out 31 Table 4 Foot Pedal Pin Ou...

Страница 6: ...y incorporated into the design of the system The lamp is mounted in an elliptical reflector with a proprietary coating to provide excellent spectral and power output Control for the OmniCure S2000 PRO...

Страница 7: ...lasses 5 Foot Pedal Switch 6 Lamp Housing Access Tool fastened beneath the system housing 7 Grounded POWER Cord 8 USB Type B Cable 9 Quick Start Guide and User Manual If your packaged unit is missing...

Страница 8: ...el Figure 1 Front Panel SYSTEM LIGHT GUIDE AND CLF STATUS INDICATOR LIGHT GUIDE PORT NFC CONTACT AREA RADIOMETER CALIBRATION PORT 5 BUTTON NAVIGATION PANEL START STOP EXPOSURE TRIGGER SD CARD SLOT 4 3...

Страница 9: ...igure 2 Rear Panel PLC CONNECTOR ANALOG JACK FOOT PEDAL AC RECEPTACLE AND DUAL FUSE HOLDER EXHAUST FAN DUCT ATTACHMENT EXHAUST FAN USB TYPE B RJ45 ETHERNET EXHAUST FAN DUCT ATTACHMENT PRODUCT LABEL EX...

Страница 10: ...product WARNING Possibly hazardous optical radiation emitted from this product Do not stare into operating lamp Caution eye damage may result from directly viewing ultraviolet light protective eye shi...

Страница 11: ...not stare directly at the light emitted from the lamp This may be harmful resulting in eye injury Always use UV protective eyewear as indicated below Additionally protect any exposed skin with approp...

Страница 12: ...g the lamp always wear protective clothing and a face mask Operate lamp only in the S2000 PRO ELITE lamp housing This prevents direct viewing of the arc and in the case of lamp bursting contains the l...

Страница 13: ...s operational life can be significantly shortened if it is handled incorrectly Do not touch the bulb s glass envelope or the inside surface of the reflector Skin oils can cause the lamp module to fai...

Страница 14: ...LITE curing system can detect and adjust its parameters for all lamp options offered with the product Please follow these steps to install the lamp 6 1 1 Be sure the AC POWER cord is disconnected from...

Страница 15: ...lamp rim fits into the mounting groove on the lamp holder assembly Figure 4 Lamp Direction 6 1 5 Make sure the middle of the lamp is in position to fit into the spring clamp Slide the lamp until it s...

Страница 16: ...mate located on the side of the lamp housing wall INTELLI LAMP CONNECTOR POWER CONNECTOR Tip the Intelli lamp connector will only attach in the correct orientation If you are having difficulty attachi...

Страница 17: ...p as required 6 1 9 Replace the lamp housing side panel and tighten the fastening screw Note 1 if the lamp housing panel is not secured completely the lamp will not strike and the system will not turn...

Страница 18: ...pull out firmly 7 1 5 Refer to section 21 5 for Light Guide Cleaning Instructions Note 3mm single leg liquid light guides are not compatible with the S2000 PRO ELITE Tip When the light guide is fully...

Страница 19: ...ure a stable optical output Warning If the lamp is turned off and an attempt is made to turn it back on before it has fully cooled the cool message will appear on the display The lamp will automatical...

Страница 20: ...brated The display will show an absolute value of irradiance in W cm2 9 1 7 Relative mode refers to the condition when S2000 PRO ELITE has not been calibrated The display will show the percentage of i...

Страница 21: ...de Additionally the current light guide diameter value selected by the radiometer will be transferred to the S2000 PRO ELITE This light guide diameter will be used by the S2000 PRO ELITE to calculate...

Страница 22: ...is accessible only when the S2000 PRO ELITE has been calibrated with an R2000 Radiometer 10 1 Adjusting the Light Output in the Relative Mode CAL LED is OFF 10 1 1 Blank Note The Closed Loop Feedback...

Страница 23: ...with the original setting 5 or 200mW cm2 whichever is greater 10 2 Adjusting the Light Output in the Absolute Mode 10 2 1 Blank Note While in Level Mode pressing the start stop button will manually op...

Страница 24: ...again to close the shutter The timer will show the total time the shutter was opened Note While in Absolute Mode and while the shutter is closed the displayed irradiance value is the user defined Set...

Страница 25: ...the system s settings can be made This can help to ensure process control when multiple operators are using the same equipment 11 1 2 To Lock the run screen press the lock button 11 1 3 To Unlock the...

Страница 26: ...Left or Right navigation buttons 12 1 3 Blank 12 2 Running a Timed Exposure 12 2 1 While in Counted Down mode press the start stop button The shutter will open and the exposure time will begin to coun...

Страница 27: ...ated from the control circuitry 13 2 Input Output Signals and Descriptions 13 2 1 Output Signals an output signal is said to be active when the output is ON a signal is said to be ON when the transist...

Страница 28: ...User Guide 27 Figure 6 Active Low Signal Circuit Configuration Figure 7 Active High Signal Circuit Configuration...

Страница 29: ...iode cathode side logic Level active low zero voltage input max open circuit output voltage 6 VDC max current sinking requirement 5mA logic low time for momentary 150mS minimum The diagram below is a...

Страница 30: ...ut current 1mA Intensity Control Input input voltage 1 5VDC maximum input current 500 A Four I O signal connectors are located on the rear panel of the S2000 PRO ELITE and one connector located is loc...

Страница 31: ...User Guide 30 13 2 6 Blank Table 2 P3 Pin Out...

Страница 32: ...3 mm audio style jack has the following pin out Connection Point SIGNAL NAME Outer Rim Positive input active low Center Pin Common ground GND Table 4 Foot Pedal Pin Out 13 4 2 This is a simple 2 wire...

Страница 33: ...User Guide 32 15 Messages Indicators The S2000 PRO ELITE display provide information to the user to aid in unit operation and to advise when certain conditions exist...

Страница 34: ...cess Green Blinking 1 second Closed loop feedback active Green Light guide not detected Red Light guide detected closed loop feedback active Green Light Guide detected closed loop feedback at 95 to 99...

Страница 35: ...system is designed to provide remote automated control of the UV spot curing system from a PC 18 1 2 The S2000 PRO ELITE PC software requirement specification below is used to describe the communicat...

Страница 36: ...sponse from S2000 PRO ELITE CLOSE 3 Lock the front panel Command to S2000 PRO ELITE LOC Response from S2000 PRO ELITE Received 4 Unlock the front panel Command to S2000 PRO ELITE ULOC Response from S2...

Страница 37: ...S2000 PRO ELITE TON Response from S2000 PRO ELITE Received 11 Turn Off Lamp Command to S2000 PRO ELITE TOF Response from S2000 PRO ELITE Received 12 Get Lamp Configuration Command to S2000 PRO ELITE...

Страница 38: ...S2000 PRO ELITE CLC Response from S2000 PRO ELITE Done 16 Get exposure time Command to S2000 PRO ELITE GTM Response from S2000 PRO ELITE n n represents a integer number in 100 s of msecs Example If 2...

Страница 39: ...r is responded no light guide diameter information is available If the unit is in LEVEL mode Need CAL will be the response If 3 403823466e38XX r is responded no light guide diameter information is av...

Страница 40: ...um irradiance output Command to S2000 PRO ELITE GMP Response from S2000 PRO ELITE n n represents a float number in Watts per square cm If the unit is in LEVEL mode Need CAL will be the response Exampl...

Страница 41: ...and n is a float number Example If SPW2 55XX r is set the S2000 PRO ELITE LEVEL set point will be 2 55 W Response from S2000 PRO ELITE Received if the value n is valid Invalid if the value n is out of...

Страница 42: ...5 Command Timing Specification 18 5 1 The minimum amount of time required for data to be received by the PC is over 20mS except for any commands related to iris movement which is over 200msec The nex...

Страница 43: ...PutString Textcom send the command GLG ComPort PutChar Asc Command_CRC 4 send the CRC8 in hex string ComPort PutChar Asc Command_CRC 0xF send the CRC8 in hex string ComPort PutChar r a 0 UnitResponse...

Страница 44: ...ocessMessages return if TimeOut MessageDlg No response from Unit mtError TMsgDlgButtons mbCancel NULL Screen Cursor crDefault Cmd33 Enabled true Timer1 Enabled true Application ProcessMessages Bit 0 1...

Страница 45: ...for i 0 i 8 i data loop A Data get first data byte for LoopCntr 0 LoopCntr 8 LoopCntr A 1 8 bit loop if A CRC8 0x01 test bit 0 of OneWire Data XOR CRC8 CRC8 0x18 toggle bits 3 and 4 of CRC8 CRC8 1 ro...

Страница 46: ...place the air filter found under the front face plate frequently to ensure unrestricted air flow It is recommended as a minimum that the air filter be removed and washed with a mild detergent and wate...

Страница 47: ...ule on the rear panel 21 2 2 Turn off the main POWER switch and remove the AC POWER cord from the unit 21 2 3 Gently pull out the drawer with the aid of a flat head screwdriver 21 2 4 Carefully lever...

Страница 48: ...system will be identified on the packaging label as well as on the filter assembly 21 5 Light Guide Cleaning Instructions for cleaning the light emitting end output of the light guide Materials Lens T...

Страница 49: ...es 2 If required remove light guide output end from the production fixture 3 Remove the light guide from the light guide port of the UV spot curing system 4 Place the protective cap onto the light gui...

Страница 50: ...49 22 Troubleshooting Display Decimal Point Status Alarm Message Description Table 9 Message References Note The Alarm will be On if the light guide is not fully inserted and the shutter is supposed...

Страница 51: ...User Guide 50 LED Indicator Status Description Table 10 Front Panel LED Descriptions...

Страница 52: ...ecurely plugged into the AC inlet on the rear of the unit 3 The mains AC POWER switch is in the ON position 22 1 2 If the LED display lights and the fan starts but the lamp will not turn on check for...

Страница 53: ...ales representative for information on purchasing a new lamp or light guide 22 1 5 If the LCD display does not light 1 If the fan is functional POWER down the unit wait approximately 20 seconds then P...

Страница 54: ...xposure time setting whichever is greater OmniCure S2000 PRO ELITE Output Accuracy with close loop feedback activated 5 or 200mW cm2 whichever is greater OmniCure S2000 PRO ELITE Minimum Adjustable Ir...

Страница 55: ...bulb 60VDC nominal Lamp Voltage range 36VDC to 95VDC Lamp Current 3 33A typical 5 7A maximum Warning The method in which lamps are disposed of must comply with local rules regulations for disposal of...

Страница 56: ...240VAC 50 60Hz Current 3 5A max at 120VAC 2 0A max at 240VAC Input Surge 50A max cold start Protection Short circuit auto recovery Overvoltage up to 135 5 of nominal EMI filtering integrated into the...

Страница 57: ...bed in more detail in section 19 The 3rd and 4th ports 15 pin DE style are for PLC interfacing A 5thport is a RS 232 port located on the side of the front panel which is used to connect to the R2000 R...

Страница 58: ...lamps switch mode POWER supplies pulsed lasers x ray equipment welding equipment and RFI generators are a few classic examples of large noise producing equipment 23 6 2 Determining the Noise Level in...

Страница 59: ...ble ferrite shield beads are P N 0443164251 by Fair Rite Products Corp or P N 28A2025 0A0 by Steward Noise suppression products are also available in the form of AC power bars that incorporate surge s...

Страница 60: ...is equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operat...

Страница 61: ...pareil doit accepter tout brouillage radio lectrique subi m me si le brouillage est susceptible d en compromettre le fonctionnement WARNING Changes or modifications not expressly approved by Excelitas...

Страница 62: ...tact your local or regional waste administration 24 5 China RoHS The symbol above indicates that this product is in compliance with China RoHS requirements 24 6 Mechanical Specifications Dimensions He...

Страница 63: ...847 5 kHz Subcarrier Data ISO_14443 Type A Modified Manchester ISO_14443 Type B NRZ_L BPSK 24 8 Miscellaneous Display 4 3 LCD touch screen Keypad 6 tactile buttons 4 navigation 1 select 1 start stop S...

Страница 64: ...reporting to a an authorized Excelitas Canada Service Centre Excelitas Canada will repair or replace these reported defects free of charge The equipment must be sent postage and carriage paid 25 1 5...

Страница 65: ...repairs not carried out by an Excelitas Canada Service Centre to the S2000 PRO ELITE system the guarantee ceases to be valid 25 3 Returning your S2000 PRO ELITE to Excelitas Canada 25 3 1 Please make...

Страница 66: ...668 8752 USA and Canada Fax 1 905 821 2055 https www excelitas com omnicure x cite inquiries Technical Assistance techsupport excelitas com https www excelitas com ox_service_request_form For a compl...

Страница 67: ...User Guide 66 27 Addendums 27 1 1 Blank...

Отзывы: