User Guide
- 46
–
21.1 Replacing the Lamp Module
21.1.1
The S2000 PRO/ELITE comes with a choice of a standard or surface curing
lamp. Please ensure that the appropriate type lamp has been selected for your
curing application. The part number of the lamp module is available on the
lamp box.
Excelitas Canada Part #
Description
012-*****
Standard Curing Lamp
012-*****
Surface Curing Lamp
Table 7 Lamp Part Number
Refer to Section 6 for lamp module replacement instructions
Refer to Section
for messages regarding troubleshooting and lamp module
replacement.
21.2 Replacing the External Fuses
21.2.1
The external (mains) fuses are located in the fuse drawer which is located in
the AC inlet module on the rear panel.
21.2.2
Turn off the main POWER switch and remove the AC POWER cord from the
unit.
21.2.3
Gently pull out the drawer with the aid of a flat-head screwdriver.
21.2.4
Carefully lever one end of the blown fuse up from its retaining clip with a small
flat-head screwdriver and lift it out.
21.2.5
Replace the damaged fuse(s) only with the same type and rating (F5A, 250V).
The rear compartment must contain two active fuses.
21.2.6
Close the fuse drawer.
21.2.7
Reconnect the AC POWER cord.
21.3 Replacing the Air Filter
21.3.1
The external air filter is located under the front face plate of the S2000
PRO/ELITE.
21.3.2
Turn off the main POWER switch and remove the AC POWER cord from the
unit.
21.3.3
Remove screws holding the filter cover in place and remove the filter.
21.3.4
Push in the replacement filter so that it sits flat in place and re-install filter
cover.