User Guide
- 18
–
8 Powering Up and Powering Down
8.1.1
Ensure that the lamp and light guide have been properly installed and that the
lamp housing panel is securely fastened.
8.1.2
Plug the S2000 PRO/ELITE unit into a properly grounded AC outlet.
8.1.3
Turn on the mains POWER switch “I “, located on the rear panel and check the
fan for airflow.
8.1.4
As soon as the LED screen turns on, a splash screen will appear and transition
into a warm-up indicator screen.
8.1.5
Phase 2 (warmp-up) takes approximately 4 minutes. Once the timer has run
out, you may enter the run screen.
Excelitas Canada
recommends 20 minutes
to ensure a stable optical output
.
Warning: If the lamp is turned off, and an attempt is made to turn it back on
before it has fully
cooled, the “cool” message will appear on the
display. The lamp will automatically re-strike when the lamp has
cooled.
8.1.6
To power down the unit, set the mains power located on the back
panel to “0”.
8.1.7
10 exposures must be run to save any new exposure settings (intensity &
timer) into memory for it to remember next time the system is powered-up.
Lamp Warm-UP:
The ARC lamp has 3 distinct phases of operation;
1.Ignition.
2
.
Warm-up.
Excelitas Canada
recommends 20 minutes of proper warm-up
and to ensure a stable optical output.
3. Stable Operation.
It is recommended that phase 1 and 2 are not interrupted. This can result in shortened
lamp life.
The lamp must be allowed to warm-up uninterrupted.