FILL, DIP, READ
NITRATE
REMOVE STRIP
Remove one eXact
®
Strip Micro NO
3
Nitrate and set in a
dry, convenient place. Replace cap on bottle.
5
READ RESULTS
READ
result displayed as Total Nitrate (as NO
3
). This
result is automatically stored in the No3 menu. After
testing is complete, rinse the sample cell immediately
and clean with brush.
7
DIP STRIP AND PRESS READ
Press
to initiate a 20 second countdown and
simultaneously
DIP
the eXact
®
Strip by submerging all
pads in the sample then use a gentle, constant, back
and forth motion (2 strokes per second) until the timer
displays
‘1’
. Remove and discard the strip. The display
will automatically start to count up for 600 seconds.
6