11
Sensafe.com
FILL, DIP, READ
REMOVE STRIP
Remove one eXact
®
Strip Micro AL Total Alkalinity
and set in a dry, convenient place. Replace cap
on bottle.
5
READ RESULTS
READ
result displayed as Total Alkalinity. This result is
automatically stored in the AL4 menu. After testing is
complete, rinse the sample cell immediately.
7
DIP STRIP AND PRESS READ
Press
to initiate a 20 second countdown and
simultaneously
DIP
the eXact
®
Strip by submerging all
pads in the sample then use a gentle, constant, back
and forth motion (2 strokes per second) until the timer
displays
‘1’
. Remove and discard the strip.
6
TOTAL ALKALINITY
If result reads "HI", repeat steps 5-7 with a Total Alkalinity Range
Extender Strip (part no. 486665) until a numerical value is
displayed; each strip adds 130 ppm to the
fi
nal result.
TIP