TROUBLESHOOTING

Listed below are possible situations that may arise while testing. 

Please contact one of our knowledgeable customer service 

representatives if you require further assistance. 

Subject

Cause

Solution

Dim screen or no 

response from meter

Low battery

Replace batteries

“LO” on LCD while 

zeroing

Low battery

Replace batteries

Dirty cell

Clean cell

Cloudy sample

Dilute sample or use 

fi

 lter

Bad LED

Contact ITS

“HI” on LCD while 

reading 

Result above 

detection level

Re-run test to verify result

“LO” on LCD while 

reading 

Result below 

detection level

Re-run test to verify result

WELL EXACT

®

 EZ ACCURACY

All tests have been calibrated using certi

fi

 ed reference standards 

and analytical spectrophotometric methods. The Well eXact

®

 

EZ has been factory calibrated and will stay valid because of its 
exceptional quality. We are so con

fi

 dent in the Well eXact

®

 EZ, we 

offer an industry leading 2-year warranty. 

We built the Well eXact

®

 EZ to be easy, accurate and 

environmentally friendly. We have achieved this by utilizing our 
patented eXact

®

 Strip Micro Technology, which uses 60% less 

water and chemistry than alternative methods. Instead of using 
a 10mL water sample, eXact

®

 Strip Micro uses a 4mL water 

sample. The accuracy of the meter is maintained by designing 
the sample cell with an 11mm path-length.

Ready Snap

®

 3 is a calibration veri

fi

 cation 

solution with predetermined color to verify the 

accuracy of your Well eXact

®

 EZ photometer. 

The easy 3 step procedure (snap, 

fi

 ll, and 

test) allows for quick veri

fi

 cation of test 

parameters. Each box contains 10 ampoules 

of 10mL solution with no dilution necessary.

METHOD VERIFICATION

WELL EXACT

®

 EZ ASSIGNED VALUE FOR READY SNAP

® 

3

READY SNAP

® 

3

TOTAL HARDNESS 

DESIRED VALUE

TOTAL HARDNESS

ACCEPTABLE RANGE

Red dye # 505

426 ppm

358 - 494

Note: 

Values indicated are speci

fi

 c to Well eXact

®

 EZ readings and current 

concentrations as found at time of manufacture.

Содержание Well EZ

Страница 1: ...ADVANCED PHOTOMETER SYSTEM Well DIGITAL ACCURACY FOR RELIABLE WATER CARE QUICK START GUIDE...

Страница 2: ...T EZ PHOTOMETER Thank you for your Well eXact EZ purchase This guide will quickly walk you through the technical details of your new photometer After initial set up test procedures and tips you will b...

Страница 3: ...sure not to over tighten SAMPLE CELL 4 ML Built in plastic 4mL cell for collecting your water sample LCD DISPLAY Displays results and test name LIGHT SOURCE Uses 525nm wavelength and 11mm path length...

Страница 4: ...ith the water sample to be tested and FILL to the top to begin test 2 In bright light situations use cell cover for zeroing and reading TIP FILL DIP READ 3 SELECT TEST Press and re press the button un...

Страница 5: ...sult is automatically stored in the PH1 menu After testing is complete rinse the sample cell immediately 7 DIP STRIP AND PRESS READ Press to initiate a 20 second countdown and simultaneously DIP the e...

Страница 6: ...econds Ignore result displayed Clean cell with Distilled Vinegar 5 or 0 1N HCl before filling the meter with the sample to be tested If running multiple tests in a row using the same water sample the...

Страница 7: ...AD Press to initiate a 20 second countdown and simultaneously DIP the eXact Strip by submerging all pads in the sample then use a gentle constant back and forth motion 2 strokes per second until the t...

Страница 8: ...the water sample to be tested and FILL to the top to begin test 2 3 SELECT TEST Press and re press the button until the display shows No3 ZERO METER Press and the LCD Display reads 0 00PPM indicating...

Страница 9: ...ored in the No3 menu After testing is complete rinse the sample cell immediately and clean with brush 7 DIP STRIP AND PRESS READ Press to initiate a 20 second countdown and simultaneously DIP the eXac...

Страница 10: ...ve the most accurate result TIP FILL CELL Rinse the cell 3 times with the water sample to be tested and FILL to the top to begin test 2 ZERO METER Press and the LCD Display reads 0PPM indicating the m...

Страница 11: ...sample cell immediately 7 DIP STRIP AND PRESS READ Press to initiate a 20 second countdown and simultaneously DIP the eXact Strip by submerging all pads in the sample then use a gentle constant back a...

Страница 12: ...water sample to be tested and FILL to the top to begin test 2 1 POWER ON PHOTOMETER Press the button to power on the Well eXact EZ 3 SELECT TEST Press and re press the button until the display shows T...

Страница 13: ...ys 1 Remove and discard the strip 6 READ RESULTS READ result displayed as Total Hardness This result is automatically stored in the TH5 menu After testing is complete rinse the sample cell immediately...

Страница 14: ...iceably smaller or larger in width than the normal strips in the bottle These should be discarded Using these strips may give unreliable results Meter is not compatible for use with powder pillows tab...

Страница 15: ...t in the Well eXact EZ we offer an industry leading 2 year warranty We built the Well eXact EZ to be easy accurate and environmentally friendly We have achieved this by utilizing our patented eXact St...

Страница 16: ...faulty or otherwise defective contact ITS by phone 1 803 329 9712 Ext 0 or email its sensafe com to describe the problem and obtain a return authorization form before returning the photometer to ITS D...

Страница 17: ...pain may occur There may be vomiting Inhalation Absorption through the lungs can occur causing symptoms similar to those of ingestion Section 4 First Aid Measures Action Skin Contact Wash immediately...

Страница 18: ...ns Conditions to avoid Heat Materials to avoid Oxidising agents and acids Haz decomp products In combustion emits toxic fumes Section 11 Toxicological Information Engineering Measures EZ Open Reducer...

Страница 19: ...latory information given above only indicates the principal regulations specifically applicable to the product described in the material safety data sheet The user s attention is drawn to the possible...

Страница 20: ...up to date product information SENSAFE ITSSENSAFE Industrial Test Systems Inc 1875 Langston Street Rock Hill SC 29730 1 800 861 9712 Inside the U S A 1 803 329 9712 Outside the U S A Fax 1 803 329 974...

Отзывы: