COMPLIANCE TESTING

This DPD test system for Chlorine is accepted 
for reporting by most health departments 
because the tests are USEPA (DIN Standard 
38 408 G4/G5, ISO 7393/2) accepted for 
testing requirements for Free Chlorine and Total Chlorine.

The compliance requirement is a photometer wavelength to 
measure between 490 and 530nm. The Pool eXact

®

 EZ smart 

photometer uses a 525nm wavelength and 11 mm path-length. 
The eXact

®

 Strip Micro CL (DPD-1) use the same reagents and 

proportions, and the resulting solution pH is maintained between 
6.2 and 6.5 as speci

fi

 ed by AWWA method 4500-Cl G/ClO2-D. 

The USEPA does not “approve” commercial DPD delivery 
systems. The eXact

®

 Strip Micro CL (DPD-1) for

 Free Chlorine

and the eXact

®

 Strip Micro CL (DPD-3) for 

Combined Chlorine

and the eXact

®

 Strip Micro CL (DPD-4) for 

Total Chlorine

 meet 

your reportable testing requirements because the eXact

®

 Strip 

Micro strips deliver the same chemicals in identical proportions. 
Consult with your local health department for of

fi

 cial regulations.

COMPONENT (FREE CHLORINE) AWWA 4500-CL G

POOL EXACT

®

 EZ

Anhydrous DPD sulfate

1.5%

1.5%

Anhydrous Na

2

HPO

4

33.4%

33.4%

Anhydrous KH

2

PO

4

Na

2

64.0%

64.0%

EDTA 1.1%

1.1%

ABOUT

ITEM

INCLUDES

PART NO.

Standard 

Carrying case

Blue carrying case with foam inserts 
(holds up to 7 bottles)

486111

XL 

Carrying case

Black carrying case with foam inserts
(holds up to 14 bottles)

486001

Made of sturdy material lined with 

foam, the carrying case offers storage 

for a Pool eXact

®

 EZ photometer and 

eXact

®

 Micro reagents.

CARRYING CASE FOR POOL EXACT

®

 EZ

Store all your necessary reagents together with your 

Pool eXact

®

 EZ in our convenient carrying case!

TIP

 

Содержание Pool EZ

Страница 1: ...1 Sensafe com QUICK START GUIDE Pool ADVANCED PHOTOMETER SYSTEM DIGITAL ACCURACY FOR BETTER POOL CARE...

Страница 2: ...inity Calcium Hardness Phosphate Chloride and Copper WHAT YOU WILL NEED TO GET STARTED Four 4 AAA batteries 4 Phillips head screwdriver Note This system has been calibrated for use with only our eXact...

Страница 3: ...meter on and creates a baseline for your water testing CELL COVER Covers the cell for mixing and bright light situations Protected against the effects of immersion in water to depth between 15cm and 1...

Страница 4: ...R Press and the LCD Display reads 0PPM indicating the meter is ready for testing 4 FILL DIP READ In bright light situations use cell cover for zeroing and reading TIP FILL CELL Rinse the cell 3 times...

Страница 5: ...initiate a 20 second countdown and simultaneously DIP the eXact Strip by submerging all pads in the sample then use a gentle constant back and forth motion 2 strokes per second until the timer display...

Страница 6: ...Pool eXact EZ 3 SELECT TEST Press and re press the button until the display shows PH2 ZERO METER Press and the LCD Display reads 0 0PH indicating the meter is ready for testing 4 In bright light situ...

Страница 7: ...result is automatically stored in the PH2 menu After testing is complete rinse the sample cell immediately 7 DIP STRIP AND PRESS READ Press to initiate a 20 second countdown and simultaneously DIP th...

Страница 8: ...ows CL3 ZERO METER Press and the LCD Display reads 0 00PPM indicating the meter is ready for testing 4 In bright light situations use cell cover for zeroing and reading TIP Use the included cleaning b...

Страница 9: ...nstant back and forth motion 2 strokes per second until the timer displays 1 Remove and discard the strip 6 READ RESULTS READ result displayed as Free Chlorine for result as Bromine multply value by 2...

Страница 10: ...dure is a continuation from Free Chlorine Test Method on previous pages DO NOT DISCARD SAMPLE FROM FREE CHLORINE TEST REMOVE STRIP Remove one eXact Strip Micro CL DPD 3 Combined Chlorine and set in a...

Страница 11: ...lorine reaction at all levels Chlorine Dioxide ClO2 Color similar to free chlorine reaction at all levels Copper Cu 2 Color development is reduced above 10 ppm mg L Iodine I2 Color similar to free chl...

Страница 12: ...g the meter is ready for testing 4 In bright light situations use cell cover for zeroing and reading TIP Clean CELL with 0 1N HCl Distilled Vinegar 5 or Muriatic Acid before filling the meter with the...

Страница 13: ...IP the eXact Strip by submerging all pads in the sample then use a gentle constant back and forth motion 2 strokes per second until the timer displays 1 Remove and discard the strip The display will a...

Страница 14: ...R Press and the LCD Display reads 0PPM indicating the meter is ready for testing 4 FILL DIP READ In bright light situations use cell cover for zeroing and reading TIP FILL CELL Rinse the cell 3 times...

Страница 15: ...s CaCO3 This result is automatically stored in the CA5 menu After testing is complete rinse the sample cell immediately 7 FILL DIP READ DIP STRIP AND PRESS READ Press to initiate a 20 second countdown...

Страница 16: ...ZERO METER Place cell cover on sample cell before zeroing Press and the LCD Display reads 0PPM indicating the meter is ready for testing 4 FILL DIP READ FILL CELL Rinse the cell 3 times with the wate...

Страница 17: ...READ result displayed as Cyanuric Acid This result is automatically stored in the CY6 menu After testing is complete rinse the sample cell immediately 7 FILL DIP READ CAP PRESS READ Place the cell cov...

Страница 18: ...ERO METER Press and the LCD Display reads 0 00PPM indicating the meter is ready for testing 4 In bright light situations use cell cover for zeroing and reading TIP FILL CELL Rinse the cell 3 times wit...

Страница 19: ...the eXact Strip by submerging all pads in the sample then use a gentle constant back and forth motion 2 strokes per second until the timer displays 1 Remove and discard the strip The display will aut...

Страница 20: ...the 2 0mL line very precisely plunger ring should line up at the 2 0mL line and little or no air bubble should be present 3 Add the syringe content 2 0mL salt water sample to a clean 50mL graduated c...

Страница 21: ...and the LCD Display reads 0 00PPM indicating the meter is ready for testing 4 In bright light situations use cell cover for zeroing and reading TIP FILL CELL Rinse the cell 3 times with the water sam...

Страница 22: ...ubmerging all pads in the sample then use a gentle constant back and forth motion 2 strokes per second until the timer displays 1 Remove and discard the strip 6 READ RESULTS READ result displayed as C...

Страница 23: ...e Test immediately after filling the cell with the water sample To obtain optimal accuracy when testing outdoors sunlight use the cell cover when zeroing and reading the sample Due to the strip slitti...

Страница 24: ...technical information and how to videos For additional technical support call 803 329 9712 Industrial Test Systems Inc 1875 Langston Street Rock Hill SC 29730 800 861 9712 or 803 329 9712 its sensafe...

Страница 25: ...email its sensafe com to describe the problem and obtain a return authorization form before returning the photometer to ITS Damage caused by improper packing of the photometer for return shipment to I...

Страница 26: ...hlorine and the eXact Strip Micro CL DPD 3 for Combined Chlorine and the eXact Strip Micro CL DPD 4 for Total Chlorine meet your reportable testing requirements because the eXact Strip Micro strips de...

Страница 27: ...27 Sensafe com NOTES...

Страница 28: ...rial Test Systems Inc Rock Hill SC USA EXACT MICRO TESTS REAGENTS PARAMETER TEST 1 PART RANGE ppm BEST ACCURACY OF TESTS Alkalinity Total 486641 10 200 10 100 Alkalinity Total Range Extender 2 486665...

Отзывы: