2016, Industrial Test Systems, Inc. Printed in USA

WELCOME TO YOUR NEW POOL EXACT

®

 EZ 

PHOTOMETER.

Thank you for your Pool eXact

®

 EZ purchase! This guide will quickly 

walk you through the technical details of your new photometer. After 
initial set-up, test procedures, and tips, you will be on your way to 
digital water testing! Each test will require the use of one or more of 
the testing methods outlined in this manual.

YOUR POOL EXACT

®

 EZ COMES WITH:

 

Cleaning Brush

 

Quick Start Guide (this booklet)

 

48 foil packet strips (6 of each): Free Chlorine (DPD-1), 
Combined Chlorine (DPD-3), pH, Total Alkalinity, Calcium 
Hardness, Phosphate, Chloride, and Copper

WHAT YOU WILL NEED TO GET STARTED:

 

Four (4) AAA batteries

 #4 Phillips head screwdriver

Note: 

This system has been calibrated for use with only our 

eXact

®

 Micro reagents.

CONTENTS

Welcome

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2

Pool eXact

®

 EZ Overview

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3

       Install “AAA” Batteries

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3

Fill, Dip, Read

     Total Alkalinity

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

4-5

     pH

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

6-7

     Free Chlorine

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

8-9

     Combined Chlorine

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

10-11

     eXact

®

 Strip Micro CL Interferences . . . . . . . . . . . . . . . . . . . . 11

     Phosphate  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-13

     Calcium Hardness

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

14-15

     Cyanuric Acid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16-17

     Copper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  18-19

     Chloride (Salt). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20-22

For Best Accuracy

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

23

Troubleshooting

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

24

    Pool eXact

®

 EZ Accuracy

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

24

    Technical Support

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 

24

About

     Built-in Sample Cell

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

25

     Warranty (2 years)

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

25

     Method Veri

fi

 cation Ready Snap

®

 3

. . . . . . . . . . . . . . . . . . . . .

25

     Compliance Testing (USEPA)

 . . . . . . . . . . . . . . . . . . . . . . . . .

26

     Carrying Case

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

26

Notes

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

27

eXact

® 

 Micro Tests & Reagents

 . . . . . . . . . . . . . . . . . . . . . . . .

28

Patent Information

 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

28

Содержание Pool EZ

Страница 1: ...1 Sensafe com QUICK START GUIDE Pool ADVANCED PHOTOMETER SYSTEM DIGITAL ACCURACY FOR BETTER POOL CARE...

Страница 2: ...inity Calcium Hardness Phosphate Chloride and Copper WHAT YOU WILL NEED TO GET STARTED Four 4 AAA batteries 4 Phillips head screwdriver Note This system has been calibrated for use with only our eXact...

Страница 3: ...meter on and creates a baseline for your water testing CELL COVER Covers the cell for mixing and bright light situations Protected against the effects of immersion in water to depth between 15cm and 1...

Страница 4: ...R Press and the LCD Display reads 0PPM indicating the meter is ready for testing 4 FILL DIP READ In bright light situations use cell cover for zeroing and reading TIP FILL CELL Rinse the cell 3 times...

Страница 5: ...initiate a 20 second countdown and simultaneously DIP the eXact Strip by submerging all pads in the sample then use a gentle constant back and forth motion 2 strokes per second until the timer display...

Страница 6: ...Pool eXact EZ 3 SELECT TEST Press and re press the button until the display shows PH2 ZERO METER Press and the LCD Display reads 0 0PH indicating the meter is ready for testing 4 In bright light situ...

Страница 7: ...result is automatically stored in the PH2 menu After testing is complete rinse the sample cell immediately 7 DIP STRIP AND PRESS READ Press to initiate a 20 second countdown and simultaneously DIP th...

Страница 8: ...ows CL3 ZERO METER Press and the LCD Display reads 0 00PPM indicating the meter is ready for testing 4 In bright light situations use cell cover for zeroing and reading TIP Use the included cleaning b...

Страница 9: ...nstant back and forth motion 2 strokes per second until the timer displays 1 Remove and discard the strip 6 READ RESULTS READ result displayed as Free Chlorine for result as Bromine multply value by 2...

Страница 10: ...dure is a continuation from Free Chlorine Test Method on previous pages DO NOT DISCARD SAMPLE FROM FREE CHLORINE TEST REMOVE STRIP Remove one eXact Strip Micro CL DPD 3 Combined Chlorine and set in a...

Страница 11: ...lorine reaction at all levels Chlorine Dioxide ClO2 Color similar to free chlorine reaction at all levels Copper Cu 2 Color development is reduced above 10 ppm mg L Iodine I2 Color similar to free chl...

Страница 12: ...g the meter is ready for testing 4 In bright light situations use cell cover for zeroing and reading TIP Clean CELL with 0 1N HCl Distilled Vinegar 5 or Muriatic Acid before filling the meter with the...

Страница 13: ...IP the eXact Strip by submerging all pads in the sample then use a gentle constant back and forth motion 2 strokes per second until the timer displays 1 Remove and discard the strip The display will a...

Страница 14: ...R Press and the LCD Display reads 0PPM indicating the meter is ready for testing 4 FILL DIP READ In bright light situations use cell cover for zeroing and reading TIP FILL CELL Rinse the cell 3 times...

Страница 15: ...s CaCO3 This result is automatically stored in the CA5 menu After testing is complete rinse the sample cell immediately 7 FILL DIP READ DIP STRIP AND PRESS READ Press to initiate a 20 second countdown...

Страница 16: ...ZERO METER Place cell cover on sample cell before zeroing Press and the LCD Display reads 0PPM indicating the meter is ready for testing 4 FILL DIP READ FILL CELL Rinse the cell 3 times with the wate...

Страница 17: ...READ result displayed as Cyanuric Acid This result is automatically stored in the CY6 menu After testing is complete rinse the sample cell immediately 7 FILL DIP READ CAP PRESS READ Place the cell cov...

Страница 18: ...ERO METER Press and the LCD Display reads 0 00PPM indicating the meter is ready for testing 4 In bright light situations use cell cover for zeroing and reading TIP FILL CELL Rinse the cell 3 times wit...

Страница 19: ...the eXact Strip by submerging all pads in the sample then use a gentle constant back and forth motion 2 strokes per second until the timer displays 1 Remove and discard the strip The display will aut...

Страница 20: ...the 2 0mL line very precisely plunger ring should line up at the 2 0mL line and little or no air bubble should be present 3 Add the syringe content 2 0mL salt water sample to a clean 50mL graduated c...

Страница 21: ...and the LCD Display reads 0 00PPM indicating the meter is ready for testing 4 In bright light situations use cell cover for zeroing and reading TIP FILL CELL Rinse the cell 3 times with the water sam...

Страница 22: ...ubmerging all pads in the sample then use a gentle constant back and forth motion 2 strokes per second until the timer displays 1 Remove and discard the strip 6 READ RESULTS READ result displayed as C...

Страница 23: ...e Test immediately after filling the cell with the water sample To obtain optimal accuracy when testing outdoors sunlight use the cell cover when zeroing and reading the sample Due to the strip slitti...

Страница 24: ...technical information and how to videos For additional technical support call 803 329 9712 Industrial Test Systems Inc 1875 Langston Street Rock Hill SC 29730 800 861 9712 or 803 329 9712 its sensafe...

Страница 25: ...email its sensafe com to describe the problem and obtain a return authorization form before returning the photometer to ITS Damage caused by improper packing of the photometer for return shipment to I...

Страница 26: ...hlorine and the eXact Strip Micro CL DPD 3 for Combined Chlorine and the eXact Strip Micro CL DPD 4 for Total Chlorine meet your reportable testing requirements because the eXact Strip Micro strips de...

Страница 27: ...27 Sensafe com NOTES...

Страница 28: ...rial Test Systems Inc Rock Hill SC USA EXACT MICRO TESTS REAGENTS PARAMETER TEST 1 PART RANGE ppm BEST ACCURACY OF TESTS Alkalinity Total 486641 10 200 10 100 Alkalinity Total Range Extender 2 486665...

Отзывы: